Issue 2, 1992

Atomic line profiles—their measurement and importance in analytical atomic spectroscopy. Invited lecture

Abstract

Profiles of the resonance lines of Ag, Co, Hg, Mn, Mg and Pt, emitted by hollow cathode lamps (HCLs), have been measured for a range of currents using a scanning échelle monochromator. The effect on the calibration graph of self-absorption broadening or self-reversal due to higher currents is complex and is strongly influenced by the presence of isotope and hyperfine structure. The échelle monochromator has also been used to measure absorption linewidths and shifts in the air–acetylene flame, and results obtained for Ag, Li, Mg and Mn were found to be in good agreement with calculated values reported in the literature. For Li, the use of a semiconductor laser diode instead of a hollow cathode light source is shown to overcome some of the inherent disadvantages of the HCL and to provide improved sensitivity and better linearity.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1992,7, 265-272

Atomic line profiles—their measurement and importance in analytical atomic spectroscopy. Invited lecture

P. L. Larkins, J. Anal. At. Spectrom., 1992, 7, 265 DOI: 10.1039/JA9920700265

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.