Comparative study of injection into a pneumatic nebuliser and tungsten coil electrothermal vaporisation for the determination of rare earth elements by inductively coupled plasma optical emission spectrometry
Abstract
Injection into a pneumatic nebuliser and vaporisation using a tungsten coil electrothermal vaporisation system, with a 3-kW argon-nitrogen inductively coupled plasma (ICP), are compared for the determination of the rare earth elements. The sampling efficiency and thus also the absolute power of detection of the tungsten coil ICP optical emission spectrometric (ICP-OES) technique are better by two orders of magnitude, than the injection technique. The absolute detection limits for the rare earth elements are at the pg level; for the refractory rare earth elements (Er, La, Lu and Y), they are lower than those obtained by graphite furnace atomic absorption spectrometry, whereas for the other rare earth elements (Eu, Sc, Tm and Yb), the detection limits are comparable. With injection of samples into a pneumatic nebuliser and ICP-OES, matrix effects are low and absolute amounts of the order of mg of the rare earth matrix can be tolerated, giving relative detection limits down to 1 µg g–1. The amount of rare earth matrix that can be tolerated with the tungsten-coil atomiser is two orders of magnitude lower. Thus the relatively detection limits of the two methods are of the same order, although the matrix effects are considerably higher with the tungsten coil.