Determination of lead in blood using flow injection and a nebuliser interface for flame atomic absorption spectrometry

Abstract

An improved system for the determination of lead in blood by flame atomic absorption spectrometry has been developed. The system is based on flow injection, a nebuliser interface and a computer signal evaluation system. The system improves the detection limit 12-fold compared with that obtained by unmodified instrumentation. A detection limit of 0.06 µM was obtained using a 10-year-old instrument. With a relative standard deviation of 5% at normal lead levels in blood, even small changes in the lead concentration can be determined with adequate precision and accuracy.

Article information

Article type
Paper

Analyst, 1988,113, 591-594

Determination of lead in blood using flow injection and a nebuliser interface for flame atomic absorption spectrometry

O. Nygren, C. Nilsson and A. Gustavsson, Analyst, 1988, 113, 591 DOI: 10.1039/AN9881300591

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.