Issue 8, 1984

Automatic continuous-flow determination of nitrate in raw and potable waters, rivers and sewage effluents by ultraviolet absorption spectrometry

Abstract

A method for the determination of nitrate, utilising continuous-flow analysis with a dialysis membrane and ultraviolet detection, is described. The method has a linear calibration range of 0–30 µg ml–1 of nitrate (as N) and a criterion of detection of 0.091 µg ml–1 of N.

The incorporation of a dialysis membrane minimised the effect of humic acid type substances that did not freely diffuse through the membrane, and also avoided the necessity to filter raw and waste water samples prior to measurement.

Article information

Article type
Paper

Analyst, 1984,109, 1053-1056

Automatic continuous-flow determination of nitrate in raw and potable waters, rivers and sewage effluents by ultraviolet absorption spectrometry

K. C. Thompson and M. Blankley, Analyst, 1984, 109, 1053 DOI: 10.1039/AN9840901053

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.