Direct determination of fluorine in high-purity water samples using vacuum sample chamber total reflection X-ray fluorescence spectrometry

Author affiliations

Abstract

A simple total reflection X-ray fluorescence (TXRF) method for the direct determination of trace F levels in water samples has been developed. This method involves the deposition of 5 μL aliquots of samples and calibration solutions on Si-wafer supports and measurement of their TXRF spectra using a vacuum chamber TXRF spectrometer with ultrathin window detector. The small size of the deposited sample fixed at the center of the Si-wafer support provided reproducible geometry that ensured the measured F Kα intensity was the same for different specimens containing the same amount of F and varied in proportion to the amount of F. This approach resulted in a calibration plot of F Kα intensity against the amount of F on TXRF specimens prepared using the same aliquot volumes containing different F concentrations. Fluorine in sample solutions was determined using this calibration plot. This approach avoided use of an internal standard and, thereby, F loss due to the acidic nature of internal standards while preparing the TXRF specimens by drying sample solutions on TXRF supports. This approach allowed determination of the F amount in synthetic samples and simulated water samples with an average deviation in the TXRF results of 4.1% from expected values and an RSD of 5.1% (1σ) for the F amount at ppm and sub-ppm levels. The use of Si-wafer supports helped to prepare small sample spots on TXRF supports with reproducible geometry and better detection limits. This also avoided interference from the O Kα peak of quartz supports with the F Kα peak.

Graphical abstract: Direct determination of fluorine in high-purity water samples using vacuum sample chamber total reflection X-ray fluorescence spectrometry

Back to tab navigation

Supplementary files

Publication details

The article was received on 17 Nov 2017, accepted on 15 Mar 2018 and first published on 16 Mar 2018


Article type: Technical Note
DOI: 10.1039/C7JA00377C
Citation: J. Anal. At. Spectrom., 2018, Advance Article
  •   Request permissions

    Direct determination of fluorine in high-purity water samples using vacuum sample chamber total reflection X-ray fluorescence spectrometry

    K. Sanyal and N. L. Misra, J. Anal. At. Spectrom., 2018, Advance Article , DOI: 10.1039/C7JA00377C

Search articles by author

Spotlight

Advertisements