Standardless semi-quantitative analysis by PIXE

Author affiliations

Abstract

A method based on the refinement of atomic and experimental parameters developed for the description of PIXE spectra is presented and applied to standardless semi-quantitative PIXE analysis. This method was implemented in the sofware PAMPA (Parameter Assessment Method for PIXE Analysis) and consists in minimizing the quadratic differences between an experimental spectrum and an analytical function proposed to describe it. The first results of PAMPA are presented for the quantification of synthetic and mineral, thin and bulk samples, and they are compared with the results obtained with a commercial software.

Graphical abstract: Standardless semi-quantitative analysis by PIXE

Back to tab navigation
Please wait while Download options loads

Publication details

The article was received on 21 Feb 2017, accepted on 11 Apr 2017 and first published on 12 Apr 2017


Article type: Paper
DOI: 10.1039/C7JA00068E
Citation: J. Anal. At. Spectrom., 2017, Advance Article
  •   Request permissions

    Standardless semi-quantitative analysis by PIXE

    T. Rodríguez, S. Limandri, S. Suárez, I. Ortega-Feliu and J. Trincavelli, J. Anal. At. Spectrom., 2017, Advance Article , DOI: 10.1039/C7JA00068E