Uranium isotope analysis by MC-ICP-MS in sub-ng sized samples

Author affiliations

Abstract

This study describes the multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) method for the determination of n(233U)/n(238U), n(234U)/n(238U), n(235U)/n(238U), and n(236U)/n(238U) isotope ratios in purified uranium solutions. The used Neptune Plus™ MC-ICP-MS is equipped with the so-called “Jet Interface”, which comprises a high capacity interface pump and specially designed sample and skimmer cones. Together with a high-efficiency sample introduction system (Aridus II™, Cetac Technologies, NE, USA), absolute U registration efficiencies of typically 1.5% to 2.5% can be achieved under routine operation. A mass bias correction, background/memory effect correction, hydride formation correction, yield and dead time correction of the multi-collector ion counters, and peak-tailing correction are performed for isotope ratio measurements. The effect of potential molecular interferences is assessed by the screening of concentrations of interfering elements in uranium fractions. The new MC-ICP-MS makes possible the determination of n(235U)/n(238U) with relative expanded uncertainties (k = 2) lower than 0.2% and allows achieving detection limits of 1 × 10−7 for n(233U)/n(238U) and 6 × 10−8 for n(236U)/n(238U) in solution aliquots containing 0.4 ng to 0.6 ng uranium. The relative expanded uncertainties (k = 2) of n(234U)/n(238U) and n(236U)/n(238U) ratios (>10−5) are below 1% and the absolute accuracy for minor isotope ratios n(234U)/n(238U) and n(236U)/n(238U) measured in CRMs is below the 1 ppm level. The developed method has been validated by comparative measurements with the previously used thermal ionisation mass spectrometry (TIMS) method by analysing test samples from an inter-laboratory comparison exercise. The MC-ICP-MS method has been applied for analysis of U isotope ratios in U fractions obtained from cotton swipe samples.

Graphical abstract: Uranium isotope analysis by MC-ICP-MS in sub-ng sized samples

Back to tab navigation

Supplementary files

Publication details

The article was received on 04 Jul 2016, accepted on 22 Sep 2016 and first published on 05 Oct 2016


Article type: Paper
DOI: 10.1039/C6JA00238B
Citation: J. Anal. At. Spectrom., 2016,31, 2272-2284
  •   Request permissions

    Uranium isotope analysis by MC-ICP-MS in sub-ng sized samples

    S. F. Boulyga, A. Koepf, S. Konegger-Kappel, Z. Macsik and G. Stadelmann, J. Anal. At. Spectrom., 2016, 31, 2272
    DOI: 10.1039/C6JA00238B