The performance of single and multi-collector ICP-MS instruments for fast and reliable 34S/32S isotope ratio measurements

Author affiliations

Abstract

The performance and validation characteristics of different single collector inductively coupled plasma mass spectrometers based on different technical principles (ICP-SFMS, ICP-QMS in reaction and collision modes, and ICP-MS/MS) were evaluated in comparison to the performance of MC ICP-MS for fast and reliable S isotope ratio measurements. The validation included the determination of LOD, BEC, measurement repeatability, within-lab reproducibility and deviation from certified values as well as a study on instrumental isotopic fractionation (IIF) and the calculation of the combined standard measurement uncertainty. Different approaches of correction for IIF applying external intra-elemental IIF correction (aka standard-sample bracketing) using certified S reference materials and internal inter-elemental IIF (aka internal standardization) correction using Si isotope ratios in MC ICP-MS are explained and compared. The resulting combined standard uncertainties of examined ICP-QMS systems were not better than 0.3–0.5% (uc,rel), which is in general insufficient to differentiate natural S isotope variations. Although the performance of the single collector ICP-SFMS is better (single measurement uc,rel = 0.08%), the measurement reproducibility (>0.2%) is the major limit of this system and leaves room for improvement. MC ICP-MS operated in the edge mass resolution mode, applying bracketing for correction of IIF, provided isotope ratio values with the highest quality (relative combined measurement uncertainty: 0.02%; deviation from the certified value: <0.002%).

Graphical abstract: The performance of single and multi-collector ICP-MS instruments for fast and reliable 34S/32S isotope ratio measurements

Back to tab navigation
Please wait while Download options loads

Supplementary files

Publication details

The article was received on 01 Aug 2016, accepted on 22 Sep 2016 and first published on 23 Sep 2016


Article type: Paper
DOI: 10.1039/C6AY02177H
Citation: Anal. Methods, 2016,8, 7661-7672
  • Open access: Creative Commons BY license
  •   Request permissions

    The performance of single and multi-collector ICP-MS instruments for fast and reliable 34S/32S isotope ratio measurements

    O. Hanousek, M. Brunner, D. Pröfrock, J. Irrgeher and T. Prohaska, Anal. Methods, 2016, 8, 7661
    DOI: 10.1039/C6AY02177H

    This article is licensed under a Creative Commons Attribution 3.0 Unported Licence. Material from this article can be used in other publications provided that the correct acknowledgement is given with the reproduced material.

    Reproduced material should be attributed as follows:

    • For reproduction of material from NJC:
      [Original citation] - Published by The Royal Society of Chemistry (RSC) on behalf of the Centre National de la Recherche Scientifique (CNRS) and the RSC.
    • For reproduction of material from PCCP:
      [Original citation] - Published by the PCCP Owner Societies.
    • For reproduction of material from PPS:
      [Original citation] - Published by The Royal Society of Chemistry (RSC) on behalf of the European Society for Photobiology, the European Photochemistry Association, and RSC.
    • For reproduction of material from all other RSC journals:
      [Original citation] - Published by The Royal Society of Chemistry.

    Information about reproducing material from RSC articles with different licences is available on our Permission Requests page.

Search articles by author