An isotopic perspective on mass bias and matrix effects in multi-collector inductively-coupled-plasma mass spectrometry

Abstract

Lead and Tl isotope intensities and ratios were measured in radial and axial plasma profiles using multi-collector inductively-coupled-plasma mass spectrometry (MC-ICP-MS). Signal profiles for 204Pb, 206Pb, 207Pb and 208Pb display mass dependent radial and axial distributions in the plasma. In radial profiles lighter isotopes exhibit greater dispersal from the plasma axis than heavier isotopes, in agreement with ICP-MS elemental observations. However, signal maxima (Imax) for heavier Pb isotopes occur closer to the load coil than Imax for lighter Pb isotopes, the reverse of ICP-MS elemental observations. These mass dependent distributions predict that the radial dispersion and axial Imax locations for 203Tl and 205Tl isotopes should occur interspersed with the values for simultaneously measured Pb isotopes. Instead however, their values are appropriate for masses greater that 208Pb, indicating that the observed profiles are not the product of a single stage mass dependent process and that other element properties also effect the distributions. Isotope ratio (and thus mass bias) variation with sampling depth reflects both the mass dependent separation of the numerator and denominator isotope Imax along the axis of the plasma and the distance along the axis of the plasma over which the isotope signals are distributed. Addition of matrix to the plasma significantly reduces the variability of these two characteristics suggesting that more accurate correction of instrumental mass bias and more accurate normalization of sample ratios to standard values could be achieved by addition of a common matrix to samples and standards.

Graphical abstract: An isotopic perspective on mass bias and matrix effects in multi-collector inductively-coupled-plasma mass spectrometry

Supplementary files

Article information

Article type
Paper
Submitted
22 Dec 2011
Accepted
17 Feb 2012
First published
20 Feb 2012

J. Anal. At. Spectrom., 2012,27, 653-662

An isotopic perspective on mass bias and matrix effects in multi-collector inductively-coupled-plasma mass spectrometry

J. Barling and D. Weis, J. Anal. At. Spectrom., 2012, 27, 653 DOI: 10.1039/C2JA10382F

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements