Linking low dispersion liquid chromatography with diode-array detection for the sensitive and selective determination of vitamins A, D and E

Abstract

A low dispersion LC system using narrow-bore column technology was linked to a diode-array detector and applied to the determination of vitamins. The LC system had been evaluated for dispersion effects. The technique was robust and reliable and proved to be extremely sensitive and selective for the determination of vitamins A, D and E. Vitamin E was assayed in various vegetable oils and the percentage distribution examined. The high mass sensitivity was demonstrated with an on-column loading of 10 ng of vitamin E, which could be clearly detected and identified.

Article information

Article type
Paper

Analyst, 1986,111, 601-604

Linking low dispersion liquid chromatography with diode-array detection for the sensitive and selective determination of vitamins A, D and E

M. Mulholland, Analyst, 1986, 111, 601 DOI: 10.1039/AN9861100601

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.