We describe the fabrication and characterisation of a capacitive pressure sensor formed by an ultra-thin graphene–polymer heterostructure membrane spanning a large array of micro-cavities each up to 30 μm in diameter with 100% yield. Sensors covering an area of just 1 mm2 show reproducible pressure transduction under static and dynamic loading up to pressures of 250 kPa. The measured capacitance change in response to pressure is in good agreement with calculations. Further, we demonstrate high-sensitivity pressure sensors by applying a novel strained membrane transfer and optimising the sensor architecture. This method enables suspended structures with less than 50 nm of air dielectric gap, giving a pressure sensitivity of 123 aF Pa−1 mm−2 over a pressure range of 0 to 100 kPa.
Existing capacitive pressure sensors employ either silicon-based or polymer-based suspended membranes that are on the order of microns in thickness, therefore requiring relatively large diameters of several 100s of microns and capacitor spacing below 1 μm in order to give a sufficiently high sensitivity.3,5 In view of improving the performance of these devices, the fabrication of large area membranes with a smaller air gap often results in membrane collapse driven by either capillary forces or stiction due to electrostatics or van der Waals forces during the fabrication or operation of the device.6,7 This limits current technologies from achieving higher sensitivities and reduces their reliability.
Emerging ultra-thin nanomaterials such as carbon nano-membranes, cross-linked gold nanoparticles and the family of 2-dimensional (2-d) materials have shown great potential as candidates for next generation nano- and micro-electromechanical systems (MEMS and NEMS).8–10 Amongst these, single-layer graphene is regarded as a promising material for pressure sensing due to its large-area processability by chemical vapour deposition (CVD), high conductivity, high elasticity and excellent tensile strength.11,12 Moreover, early studies on graphene have shown that when suspended across a micro-cavity etched into silicon oxide, the graphene exhibits an ultra-strong adhesion to the surface and thereby seals the micro-cavity off from the surrounding atmosphere.13,14 This property also holds for other 2-d materials, including CVD graphene, due to their atomic thickness and negligible bending rigidity.15 The opportunity of forming gas tight membranes with almost negligible bending rigidity and its potential for developing pressure sensitive devices from graphene has prompted many efforts into fabricating suspended CVD graphene devices.16–21
The most common method of forming suspended CVD graphene structures is to transfer the graphene from its parent substrate (on which it is synthesised) onto a cavity-baring substrate using a sacrificial polymer support film.22,23 However, in a manner similar to current capacitive pressure sensor devices, devices often collapse due to stiction or capillary forces that act on the graphene during the removal of the polymer support.20,21 Additional processing steps have been developed to overcome this fabrication challenge. For example, solvents are exchanged in order to reduce capillary forces,24 critical-point drying protocols are applied25 and samples are annealed after transfer.26 Alternatively, support-free transfer methods such as wet-contact printing,27 face-to-face transfer28 and transfers facilitated by self-assembled monolayers29 have also been developed. Whilst such methods increase the probability of individual graphene membranes to remain suspended after transfer, they do not provide a complete solution for a 100% yield of suspended graphene structures over a large area.
Another strategy in preventing the collapse of suspended graphene is to reinforce membranes with another material.30–32 For example, by stacking up multiple layers of CVD graphene on top of one another or by depositing carbon nanotubes on a graphene film, crack-free freestanding structures up to 1.5 mm in diameter and 5.1 nm thickness have been demonstrated.33,34 However, this approach requires many fabrication steps, is very time consuming and results in wrinkling of the suspended membranes. Most recently, graphene–polymer heterostructure membranes comprising single-layer CVD graphene and an ultra-thin polymer layer, 10s of nanometres in thickness, have been shown to provide an excellent support for facilitating suspended graphene devices with 100% yield.35 Whilst the presence of a polymer on the surface of graphene is often a disadvantage in chemical and bio-medical sensing applications (where a clean surface of graphene is crucial to providing a high sensitivity),36 as a mechanical component it provides an optimal trade-off between the elastic modulus and membrane yield.37
Further, a route towards minimising membrane stiction in NEMS switches is to reduce the gate electrode contact area by implementing 3-dimensional gate electrode geometries.38,39 For example, CVD graphene closely suspended over an electrode with a ‘line contact’ has been demonstrated with reliable radio frequency actuation at low operating voltages.7 In the present work, we fabricate, characterise and test capacitive pressure sensors comprising an array of ultra-thin graphene–polymer heterostructure membranes and demonstrate a route towards high sensitivity pressure sensors over a large pressure range using a modified transfer method and a 3-dimensional cavity structure.
Each sensor array now comprises a monolayer graphene membrane with a second polymer layer of total membrane thickness tm, suspended over an array of circular cavities each of diameter 2a and depth tgap. An additional SiO2 dielectric of thickness tox exists at the base of each cavity, beneath which lies the highly doped Si substrate which acts as the counter electrode. An array of such cavities covers an area of up to 1 mm2. Since the cavities are now sealed by the graphene membrane, an equilibrium pressure p0 exits within each cavity. When the external pressure is changed to a value p > p0 (positive pressure), the membrane deflects in to the cavity and the centre of the membrane achieves a deflection of z. The deflection of the membrane results in a change in capacitance which is measured and correlated to the change in pressure. A detailed description of the entire fabrication protocol is given in ESI Discussion 1.†
(1) |
A unit cell of a pressure sensing device is defined by two areal components; a suspended area, Asus and a surrounding substrate supported area, Asup as shown in Fig. 2b. In this analysis the suspended area is defined by a circle with radius a and the total unit cell area is defined by
Aunit = Asus + Asup = lx × ly. | (2) |
The total capacitance of a unit cell in an array is then given by
(3) |
(4) |
(5) |
(6) |
A similar analysis has also been shown for square plates.41 The capacitance of an entire device with an array of N × M pressure sensing cells is thus given by summing over all of the rows i and columns j of the array given by
(7) |
(8) |
Whilst this model provides a basic description of a pressure sensor array, in practice numerous other effects play a role in determining a pressure sensor's performance. A summary of these effects can be captured by an equivalent circuit model of the pressure sensor.
The fabrication protocol outlined in Fig. 1 was used to fabricate two different designs of pressure sensor. The thickness of the polymer layer used in both of these designs was calibrated such that 100% yield of suspended membranes can be achieved in both devices, whilst maintaining a high deflection sensitivity of the membranes and thus a measureable capacitance signal. This calibration procedure is outlined our previous study on graphene–polymer membranes.35 The first design uses a thick SiO2 layer, deep cavities and relatively small membranes as shown in Fig. 3a.i. This design uses a straightforward circular cavity geometry, in order to confirm that the experimental device behaviour fits the electromechanical model. The second design uses a thin SiO2 layer, shallow cavities, relatively large membranes and micro-bumps in the cavities as shown in Fig. 3a.ii. This design is to optimise the deflection sensitivity and electromechanical coupling of the pressure sensor. The micro-bumps in the cavity act as an anti-stiction surface for the membrane as explained in a later discussion. In the first part of this work we will only be concerned with design 1 in order to experimentaly verify the electromechanical model above.
Next we cycled devices containing cavities between 20 kPa and 100 kPa with a cycling period of 30 seconds. Fig. 4c shows the reference pressure and measured capacitance of a device containing cavities over 10 cycles. The device's response to the continuous change in pressure is reproducible with a precision of 6% of the measured pressure range given by the standard deviation of the capacitance value measured during the cycling experiment. This irreproducibility is likely due to the air-flow within the pressure chamber creating movement of the graphene–polymer film's edges, thus creating irregular jumps in capacitance of the device. As the inaccuracy due to the sensors reproducibility dominates over other contributions such as the sensors linearity and hysteresis, we quote the accuracy of the sensor as 6%. Finally, we measured the capacitance of the sensor for a short pressure pulse of 6 seconds at 90 kPa in order to probe the sensor's response time. Fig. 4d shows that the measured capacitance and reference pressure signals correlate well and minimal drift is observed over the complete cycle. However, a delayed response is observed on the down-step of the pressure pulse giving a combined response time of the up and down pressure stroke of 1.8 s. We expect that the formation of ripples in the membrane has a smaller time scale than ripple relaxation. A similar effect was observed previously in micro-structured dielectric capacitive pressure sensors.44 In order to confirm that the observed capacitance change is solely due to the deflection of suspended graphene–polymer membranes and that gas concentrations are independent of the device capacitance, we also measured the samples without any cavities in different gaseous atmospheres (ESI Discussion 4†). We also note that the pressure sensor forms a low leakage capacitor with a discharge resistance of 6.5 MΩ when measured at an excitation of 1 V at 1 kHz. Assuming the discharge curve is linear this gives a discharge time of approximately 32 μs. As this time scale is an order of magnitude lower than the sampling rate of our capacitance measurements, we do not expect this to have a significant effect on our measurements.
Using the height profiles of membranes pressurised to different pressure values during micro-blister inflation tests, we were able to estimate the sensitivity of the capacitive sensor using the previously derived expression (eqn (8)). As each sensor contains several thousand membranes over a 1 mm x 1 mm square, we conducted AFM measurements at 5 randomly distributed 90 μm x 90 μm sample areas, giving an average distribution of membrane morphologies of each sensor. AFM measurements were taken immediately after pressure sensing experiments to ensure our calculation captures membranes that were definitely active during sensing. Using parameters from the table in Fig. 3e, the calculated sensitivity is 0.28 ± 0.8 aF Pa−1. The large error in our calculation is attributed to the variation in deflections of individual membranes. The significance of these variations for a capacitive pressure sensor are discussed in a later section. Further, the calculation is higher in comparison to the measured sensitivity of 0.21 ± 0.01 aF Pa−1. We attribute this difference to a combination of charge leakage through the SiO2 layer and along the surface of the sensor chip as well as due to the plate model being at its limit in the pressure used throughout our experiment.
The performance of this first demonstration of a capacitive pressure sensor remains low compared to current state of the art sensors; nonetheless this design serves to validate our experimental results with a model. For example, a recently published CMOS (complementary metal–oxide semiconductor) pressure sensor spanning an area of 0.31 mm2 as described in Table 1, has a sensitivity of 1.2 aF Pa−1.45 In the following section we outline a route towards high sensitivity pressure sensors and demonstrating an initial working prototype with a sensitivity of 5.0 aF Pa−1 over an area of just 0.04 mm2.
Metal–polymer38 | Metal–silicon30 | Graphene–polymer(this work) | |
---|---|---|---|
Membrane matetial | Parylene C/AU/parylene | SiOx/Al/SiOx | CVD graphene/PMMA |
Elastic modulus | 3.9 GPa | <100 GPa | 10.9 GPa |
Membrane thickness | 1.3 μm/200 nm/1.3 μm | 500 nm/2 μm/500 nm | 0.35 nm/140 nm |
Sensitivity at FPS | 2.2 aF Pa−1 | 1.2 aF Pa−1 | 4.2 aF Pa−1 |
Areal sensitivity (mm2) | 25 aF Pa−1 | 4 aF Pa−1 | 123 aF Pa−1 |
Full pressure scale (FPS) | 6.8 kPa | 600 kPa | 100 kPa |
(9) |
In order to further improve the pull-in instability of the graphene–polymer capacitive pressure sensor we employed two additional steps to our fabrication protocol. First, micro-wells were etched into the base of the cavities in order to reduce the silicon oxide area below the suspended membrane, thereby reducing the total Casimir force acting on the membrane. Second, we applied a strain to the graphene–polymer membrane during the membrane transfer procedure in view of stiffening the membrane and therefore increasing the effective restoring force keff of deflected membranes. In addition, we also changed the shape of cavities to a hexagonal in order achieve high packing densities whilst preventing high local stresses as found in square shaped membranes.49 A schematic, optical micrograph and AFM of this optimised design is shown in Fig. 3a–d.ii. In addition a schematic of the strained transfer and an optical micrograph of the resulting suspended membrane structure in comparison to a normal membrane transfer are shown in Fig. 5. The schematic illustrates that crumpled graphene–polymer membranes as transferred using the standard method (Fig. 1) results in excess material that can readily adhere the to the cavity (Fig. 5a.ii). When these membranes are strained, such crumples are flattened out (Fig. 5b.i) and the membrane can span large areas without stiction to the cavity (Fig. 5b.ii). The optical micrographs shown in Fig. 5a.iii and b.iii show a change in contrast between collapsed and suspended membranes respectively due to applying strain during membrane transfer. The applied strain during transfer is estimated as 0.15 ± 0.01 by Raman spectroscopy as described in ESI Discussion 5.†
The morphology of a bare micro-cavity, a suspended graphene–polymer membrane and a collapsed membrane are shown in Fig. 3a–d.ii. Taking the film thickness as 140 ± 4 nm we estimate that the graphene layer is suspended just 40 ± 4 nm above the oxide layer. Further comparison of the collapsed membrane with a cross-section of the bare micro-cavity reveals negligible adhesion to the sidewalls of the sub-cavities. This infers that the membrane is loosely bound to the bed of the micro-cavity, relying solely on the adhesion to the area remaining on the cavity bed. We were able to release collapsed membranes from the cavity base by retracting the transferred film and re-laminating the surface with applied strain as shown in the ESI Video.†
A sensor with device design 2 that was optimised for high sensitivity was tested in our pressure setup. The device performance was characterised using the same procedure as described previously apart from the maximum sensing pressure was limited to a maximum of 80 kPa in order to minimise permanent membrane stiction. Fig. 6 shows the sensitivity of a device with design 2 when cycled between 0 and 80 kPa at 30 seconds intervals in comparison to a device with design 1. A significant increase in performance is measured despite the sensing area occupying just 0.25 mm2 on the chip compared to 1 mm2 in the first device. However, cycling the sensor does not give the same degree of reproducibility in sensitivity as the first generation device. Further detail on the pressure testing data of design 2 is given in ESI Discussion 6.† The large variance in the second device's response is attributed to noise in the capacitance measurements which is likely due to charge leakage through the oxide and along the chip surface. We expect the majority contribution to originate from trapped states and mobile surface charge in the gate oxide as a result of the DRIE of sub-cavities. We support this assumption by comparison of the C–V characteristics of the two generations of devices as shown in ESI Discussion 7.† In order to reduce the noise level in future iterations we propose the use of a pinhole free dielectric or deposition of a barrier layer.
In Table 1, the performance of graphene–polymer capacitive pressure sensors are compared. Metal–polymer pressure sensors typically involve membranes ranging in diameter from 500 μm up to several millimetres with polymer thicknesses of several microns encapsulating a thin metal layer (100–200 nm). The combination of a soft membrane material and large membrane diameter gives excellent sensitivity over a small pressure range. Such sensors are typically developed for biomedical applications where small pressure fluctuations in the body are monitored and the biocompatibility of the sensor's constituent materials are crucial. Metal–silicon pressure sensors with a membrane diameter of several hundred microns, consisting of a stack of two sub-micron thick SiO2 layers encapsulating a metal layer, are of a significantly higher stiffness. In addition, CMOS processing compatibility enables membranes to be suspended in close proximity to the complimentary electrode thereby giving a large sensing range whilst maintaining high sensitivity.
Graphene–polymer pressure sensors aim to cover the entire operating range for MEMS pressure sensors, including the areas occupied by high sensitivity, low range polymer sensors and low sensitivity, high range silicon sensors as well as bridging the gaps in performance between the two. Whilst our first iteration of devices demonstrate reproducible pressure sensing using graphene–polymer stacks as the active element, the large gap between the membrane and the complimentary doped silicon sensing electrode (1430 nm) gives a relatively low sensitivity. Devices with design 2 use a much thinner oxide layer in order to improve the electromechanical coupling of the graphene–polymer membrane, thereby improving the device sensitivity. Moreover, the pressure range (80 kPa) is maintained as the membrane diameter is kept considerably small and the membrane is pre-strained during the transfer procedure. The stiffness of the membranes used in all of these devices lies between that of silicon and polymer based MEMS (10.9 GPa) and the membrane transfer technique enables the fabrication of densely packed suspended membranes on the wafer scale. However, the elastic properties of the membrane can be varied with ease by modifying either the polymer thickness or number of graphene layers.35 For example, by using an even stiffer membrane in a similar architecture, as shown by design 2, there is scope to further extend the pressure range of graphene–polymer pressure sensors whilst maintaining a high sensitivity.
In addition to improving device sensitivity and range, the graphene–polymer membrane structure aims to overcome several reliability issues faced in current MEMS and NEMS technologies.47,50 First, the use of sacrificial layers in CMOS processing puts several limitations on device architecture and material design; materials must be resistant to aggressive etchants, have sufficient stability to overcome capillary forces, and an additional sealing step is required after removal of the sacrificial etchant.51,52 By transferring the active mechanical component directly onto a pre-patterned micro-cavity in air we avoid trapping liquids that initiate membrane collapse and completely seal the micro-cavity. Moreover, CMOS compatible polymers such as parylene-C have been shown to be equally valid for this fabrication process.53 A further challenge is presented by the brittle fatigue in silicon-based membranes.54 This failure mechanism is often attributed to stress corrosion cracking, whereby stress-assisted hydrolysis of the native or deposited SiO2 layer initiates crack propagation.50 Graphene and PMMA are both known to have reversible morphological changes in response to humidity and there is no indication of crack formation or propagation on cycling the pressure sensors.55–57
Whilst we recognise that the reproducibility and hence the accuracy of the graphene–polymer sensors are not comparable to state-of-the-art devices, the aim of the sensor principle is to improve the trade-off between sensitivity and pressure range of capacitive pressure sensors. We identify inaccuracies in the pressure cycling data as engineering challenges which will be addressed by refining the sensor fabrication technique and design. For example, by pre-patterning the CVD graphene before the polymer deposition step in the fabrication of the device, the exposure of graphene edges to the surrounding atmosphere is minimised. Further, the use of an oil-fill of the pressure sensor housing can be used to isolate the active sensing area.58 Both of these techniques are expected to significantly minimise noise in the capacitance signal.
A further performance inaccuracy in these sensors is the response time. We note that a delayed response time of the graphene–polymer pressure sensor is mostly apparent in the down stroke of a pressure cycle. As the deflection mechanism is dominated by the bending of the polymer component of the graphene–polymer membrane it is most appropriate to compare this response time to that of other polymer-based pressure sensors. The typical response time of polymer-based pressure sensors is on the order of 1–10 s which is primarily attributed to the relaxation time of the polymer matrix as it reconfigures into its equilibrium state.44,59 In contrast silicon-based the pressure sensors typically have a response time on the order of 10–100 ms due to the high stiffness and the crystalline nature of the material.60 Whilst the current sensor performance may not be suitable for the measurement of short pressure changes, such as in mobile hydraulics for example,61 the sensor may be more suitable for applications with long pressure cycles, such as level measurements or blood-pressure monitoring.62 In order to improve the sensor response time we propose to further reduce the thickness of the polymer layer as well as using crystalline polymer materials such as parylene-C that exhibit significantly faster relaxation times compared to the PMMA polymer that is currently used.
With the graphene–polymer pressure sensor in its early stage of development, it remains a challenge to accomplish a large area array of suspended membranes with uniform performance characteristics. Such non-uniformities arise from a combination of membrane relaxation and gas-leakage through defects in the membranes. The effects of membrane relaxation and gas-leakage have been investigated in previous micro-blister inflation experiments on graphene–polymer membranes. This study demonstrated that graphene–parylene-C membranes of thicknesses ranging from 23 nm to 201 nm have an excellent reproducibility in the deflection of individual membranes, but show a significant variation in the deflection between different membranes across a large area array. However, with respect to the performance of the capacitive pressure sensor these variations become averaged out due to the large quantity of membranes that operate in parallel across the entire array.
Increasing the number of cavities within the device therefore provides a strategy towards improving the reproducibility of the device, however this approach is limited due to the cost of increasing the device footprint by adding more cavities. A more direct approach to solve the issue of reproducibility is to improve the control over the method of laminating the graphene–polymer film onto the cavity-baring substrate, thus regulating the mechanical properties of the individual suspended membranes. Moreover, by further reducing the thickness of the polymer layer a reduction in gas-leakage can be obtained as previously shown in the aforementioned study on graphene–parylene-C membranes.
Beyond the initial demonstration of a capacitive pressure sensor, we aim to develop more sophisticated MEMS and NEMS devices using ultra-thin graphene–polymer membranes. The excellent elasticity and high temperature compatibility of graphene makes graphene–polymer membranes attractive for pressure sensing in harsh environments, where shock and elevated temperatures would cause significant damage to tradition silicon MEMS. Whilst we note that the polymer layer is likely to be the limiting material in harsh environment applications, the large variety of polymers available as ultra-thin coatings also allows us to fabricate a range of ultra-thin membranes with properties tuned to specific applications, where the polymer layer not only acts as a mechanical reinforcement but also gives additional functionality to the membrane.8,63 In view of the latter, we also envisage the use of graphene–polymer membranes in polymer MEMS devices ranging from micron-scale pumps and valves in micro-fluidics and lab-on-chip devices to pressure sensors and actuators in biomedical applications.