A.
Klančar
,
M.
Zakotnik
,
R.
Roškar
and
J.
Trontelj
*
University of Ljubljana, Faculty of Pharmacy, Department of Biopharmaceutics and Pharmacokinetics, Askerceva 7, 1000 Ljubljana, Slovenia. E-mail: anita.klancar@ffa.uni-lj.si; maticzakotnik@gmail.com; robert.roskar@ffa.uni-lj.si; jurij.trontelj@ffa.uni-lj.si; Fax: +386 1 425 80 31; Tel: +386 1 4769 649 Tel: +386 1 4769 655 Tel: +386 1 4769 547
First published on 11th August 2017
New organic pollutants such as pharmaceuticals have been recently classified as emerging pollutants. These are chemically active compounds and may pose a risk to environmental organisms as well as to humans. Nowadays, separation and analytical techniques enable detection and quantification at extremely low concentrations. We aimed to develop and optimise a new method that would minimise (organic) solvent consumption and miniaturise and simplify sample preparation through the use of stir-bar sorptive extraction of selected pharmaceuticals, followed by liquid desorption and liquid chromatography coupled to mass spectrometry. Several conditions and parameters that affect the extraction procedure (extraction time, sample pH, sample volume, extraction temperature, and addition of salt and organic modifier) were evaluated for suitable extraction efficiency. The optimised method (sample volume, 20 mL; sample pH, 9; extraction time, 150 min; and desorption time, 15 min; in an acetonitrile and methanol [50:50 v/v] mixture at 50 °C, without any added modifier) was validated and provided recoveries above 80% for 12 analytes, and for 6 analytes they were between 45% and 65%, while for another 6 analytes the recoveries were below 25%. The method provided a wide linear range (mainly 1.25–1250 ng L−1, R2 > 0.99) with very low limits of quantification for all analytes (1.25–5.0 ng L−1). The method was used to test 15 wastewater samples and 22 out of 24 monitored pharmaceuticals were detected, generally in concentrations below 200 ng L−1.
Today, it is well recognized that the increased use of the above-mentioned compounds inevitably leads to their occurrence in the environment. However, because of the large dilution factor when they enter water bodies, their concentrations mostly reach only trace levels and their determination usually requires a pre-concentration step before separation and detection. In parallel with analytical progress, modern guidelines on green analytical chemistry have been established regarding minimising (organic) solvent consumption, miniaturization and simplification.3 As a result, in the 1990s new sample preparation techniques such as solid-phase micro-extraction (SPME) and stir-bar sorptive extraction (SBSE) became of increasing interest as alternatives to more solvent-consuming techniques such as liquid and solid-phase extractions.4
The technique of SPME established the basis for further development of equilibrium partitioning or sorptive extraction. In 1999, Baltussen et al. published the theory of SBSE.5 Briefly, the efficiency of analyte partitioning into the solid phase (polydimethylsiloxane [PDMS] is the widest commercially available) can be compared to the distribution characterised by the octanol–water partition coefficient (KPDMS/W ∼ KO/W). Thus, the polarity of the organic analyte roughly predicts the extraction efficiency. Based on experimental data, an equation was set for the theoretically predicted extraction efficiency of the analyte at equilibrium (eqn (1), where EE is the extraction efficiency). According to the equation, it is assumed that successful extractions are limited to analytes with higher KO/W (logKO/W > 2.7). In addition to the KO/W, the phase ratio β (ratio of water sample volume, Vw) and volume of PDMS (VPDMS coated on stir bar) also play key roles in the extraction efficiency (eqn (2)).
EE (%) = (KO/W)/(KO/W + β) | (1) |
β = Vw/VPDMS | (2) |
Following the publication of the advantages of higher recoveries in SBSE than in SPME, several research groups have been developing SBSE for various analyses.3,6–8 The primary concept of SBSE was first designed for connection with a thermal desorption system followed by gas chromatography coupled to mass spectrometry detection.5,9 Although several methods have been developed, only a minority use the alternative approach of solvent back-extraction or liquid desorption coupled to mass spectrometry detection.10 The latter method allows cost-effective determination of analytes and also reduces solvent consumption. The SBSE method in general reduces the amount of sample needed for the analysis. In addition, the loaded stir bars can be conveniently sent to the laboratory and can be stored for a week without any significant loss of analytes.11 Moreover, the convenience of sampling with stir bars enables extended research in large environments.
In recent years, the SBSE method has been used in environmental and food analysis and also in biomedical and life science applications (reviewed by Camino-Sánchez et al.).7 Focusing solely on pharmaceuticals, this sampling technique has been used mainly on biomedical fluids such as urine, serum, plasma and saliva.3,9,12,13 However, for the purpose of environmental monitoring of pharmaceuticals, the methodology using the generally available stationary phases for SBSE is still lacking, particularly for water bodies. Apart from the reports of hormone analytics,13–17 there have been only a few reports on SBSE determination of pharmaceuticals in the environment and all have dealt with only a small number of pharmaceuticals, mostly using a combination of thermal desorption and gas chromatography.6,9,18–21 However, there are some reports on the new in-house prepared stir-bar coatings6,19,22–24 and two new commercial polar coatings (ethylene glycol modified silicone (EG Silicone Twister®) and polyacrylate (Acrylate Twister®)).21,25
In response to the lack of methodology for routine SBSE monitoring of pharmaceuticals in wastewater samples, the aim of our research was to develop and optimise SBSE for selected pharmaceuticals, followed by liquid desorption and liquid chromatographic separation coupled to mass spectrometry detection. In comparison with existing methodologies, SBSE is suitable for less well equipped laboratories as thermal desorption is not required. Moreover, this approach does not expose analytes to extremely high temperatures and enables analysis in replicates when needed. Linearity, repeatability and limits of quantification were evaluated before testing wastewater samples. Primarily, the concentrations of selected pharmaceuticals in ten effluent wastewater samples from wastewater treatment plants (WWTPs) were determined. In cases where influent samples were available, removal efficiencies for five WWTPs were also calculated.
Reagents for preparation of standards and samples included acetonitrile (ACN), dichloromethane, formic acid (98–100%), methanol (MeOH), 2-propanol and potassium dihydrogen phosphate from Merck (Germany); ammonium formate and sodium chloride were from Sigma-Aldrich. Ultra-pure water was produced in a Millipore Milli-Q water purification system A10 Advantage (Millipore Corporation, USA). Solvents for liquid chromatography-mass spectrometry (LC-MS/MS) analyses were LC-MS grade acetonitrile ChromasolV® (Sigma-Aldrich), Milli-Q water and formic acid (98–100%) Suprapur® (Merck).
GERSTEL Twister® stir bars (GERSTEL GmbH & Co., Germany) were used in SBSE procedures. The bars were 10 mm in length and externally coated with a layer of PDMS 1 mm thick. The dimension of 10 × 1 mm corresponds approximately to 63 μL of PDMS per stir bar.
After use, the stir bars were reconditioned in a mixture of dichloromethane:MeOH (50:50 v/v) for 30 min, then in pure MeOH for 30 min, at room temperature with a stirring speed of 990 rpm. The bars were dried and stored for further extraction procedures.
The optimised method was evaluated by determining the linear range, determination coefficients, limits of quantification, imprecision and extraction efficiency (Table 3). For quantification, fluoxetine-d5 was used as the internal calibration standard. The linear range was determined in 20 mL aliquots of ultra pure water spiked at eight concentration levels from 1.25 ng L−1 to 1250 ng L−1, each in duplicate. Back calculation of the concentration and bias from nominal values was determined for each calibration point. Limits of quantification were set as the lowest standard on the calibration curve that exhibited acceptable accuracy (deviation from the nominal value ±20%) and precision (expressed as relative standard deviation%) while having the signal to noise ratio >30:1.
Analyte | Retention time (min) | MRM (m/z) | CE (eV) | Frag (V) | P | Q1 resolution | Q3 resolution | ||
---|---|---|---|---|---|---|---|---|---|
a MRM, multiple reaction monitoring; CE, collision energy; Frag, fragmentor voltage; P, polarity; Q, quadrupole; wide = 1.2 amu; widest = 2.5 amu. | |||||||||
Amitriptyline | 4.43 | 278.2 | > | 91.0 | 24 | 103 | + | Widest | Widest |
Azithromycin | 2.63 | 749.5 | > | 83.0 | 69 | 200 | + | Widest | Widest |
Bromazepam | 3.56 | 316.0 | > | 182.0 | 40 | 160 | + | Widest | Widest |
Carbamazepine | 4.13 | 237.1 | > | 193.9 | 12 | 103 | + | Widest | Widest |
Clomipramine | 5.15 | 315.2 | > | 86.1 | 16 | 106 | + | Widest | Widest |
Clonazepam | 4.81 | 316.1 | > | 270.1 | 21 | 121 | + | Widest | Widest |
Desipramine | 4.02 | 267.2 | > | 72.1 | 12 | 98 | + | Widest | Wide |
Diazepam | 6.14 | 285.1 | > | 193.0 | 32 | 159 | + | Wide | Wide |
Donepezil | 3.11 | 380.2 | > | 65.0 | 97 | 86 | + | Wide | Wide |
Escitalopram | 3.40 | 325.2 | > | 109.0 | 29 | 81 | + | Widest | Widest |
Fluoxetine | 4.76 | 310.1 | > | 148.1 | 1 | 100 | + | Widest | Widest |
Haloperidol | 3.66 | 376.2 | > | 165.0 | 21 | 126 | + | Widest | Wide |
Imatinib | 2.80 | 494.3 | > | 395.2 | 25 | 200 | + | Widest | Widest |
Loperamide | 5.84 | 477.2 | > | 266.1 | 24 | 159 | + | Widest | Wide |
Loratadine | 5.96 | 383.2 | > | 337.1 | 21 | 121 | + | Widest | Widest |
Metoprolol | 2.58 | 268.2 | > | 116.0 | 12 | 96 | + | Widest | Wide |
Promethazine | 3.74 | 285.1 | > | 86.1 | 12 | 55 | + | Widest | Wide |
Propranolol | 3.11 | 260.2 | > | 56.1 | 25 | 126 | + | Widest | Widest |
Raloxifene | 3.28 | 474.2 | > | 112.0 | 32 | 200 | + | Widest | Widest |
Risperidone | 2.88 | 411.2 | > | 191.1 | 28 | 159 | + | Widest | Wide |
Selegiline | 2.43 | 188.1 | > | 91.0 | 21 | 81 | + | Widest | Widest |
Sertraline | 4.89 | 306.1 | > | 158.9 | 24 | 60 | + | Widest | Widest |
Tramadol | 2.58 | 264.2 | > | 58.1 | 17 | 81 | + | Wide | Wide |
Triclosan | 6.80 | 286.9 | > | 35.1 | 4 | 60 | − | Widest | Widest |
Venlafaxine | 2.91 | 278.2 | > | 58.1 | 17 | 81 | + | Widest | Widest |
Verapamil | 4.39 | 455.3 | > | 150.1 | 41 | 184 | + | Widest | Widest |
Ziprasidone | 3.15 | 413.1 | > | 194.0 | 25 | 161 | + | Widest | Widest |
Fluoxetine-d 5 | 4.77 | 315.2 | > | 153.2 | 1 | 100 | + | Widest | Widest |
(3) |
All WWTPs were municipal plants and received both domestic and industrial sewage. The capacities of plants A–E were 1000, 100000, 68000, 6500 and 15500 population equivalents, respectively. Plants A and B used a secondary wastewater treatment, plants C and D a tertiary wastewater treatment and plant E a primary treatment. The second set of WWTPs I–V is described in detail in Section 3.4.2, where removal efficiency is discussed. All wastewater samples were collected in the spring of 2016 and were 24 h composite samples (time-proportional) collected in clean glass bottles, and then stored at 4 °C and analysed within 24 h.
Several conditions and parameters may affect stir-bar extraction, the most important being extraction time, sample pH, extraction temperature and the addition of NaCl and MeOH. All these parameters were evaluated and optimised to achieve acceptable extraction efficiency. Since the number of varied parameters was relatively low, one-variable-at-a-time methodology was used. Following the preliminary development and optimisation experiments, some parameters such as analyte concentration, 10 μg L−1; sample volume, 20 mL; stirring speed, 990 rpm; and stir bar dimensions, 1 × 10 mm were fixed for the optimisation tests. The extraction time was fixed at 60 min for all experiments (except for those testing the influence of the extraction time) and the temperature was fixed at 50 °C (except for testing the influence of the temperature). To facilitate comparison between parameters during the extraction processes, conditions for the desorption process were fixed as follows: desorption solvent, 2 mL ACN:MeOH (50:50 v/v); stirring for 15 min at 990 rpm; and temperature, 50 °C.
The influence of sample pH on SBSE efficiency was evaluated within the pH range of 3–9, proposed as harmless to the PDMS phase. The regulation of sample pH was assessed at four levels: pH 3, 5, 7 and 9. The relative extraction efficiency for each analyte was calculated with reference to the highest extraction efficiency (100%). As shown in Fig. 1, samples at pH 9 provided the best results for almost all analytes, except for propranolol and ziprasidone, where recovery was highest at pH 7; triclosan reached maximum at pH 5. This observation is closely in line with pH partition principles, i.e. the theory of distribution of the un-ionized form of analyte and the relationship between pKa and pH values.
After the pH was set at 9, the effect of NaCl addition was tested from 0 to 2000 mg (0–10% w/v). An increase of salt gradually decreased almost all extraction efficiencies (Fig. 2A). Such a decrease could be the result of increased viscosity with the salt addition, as this could slow down the extraction kinetics of these analytes or affect the extraction polymer sorbent. Similar results have been reported.6,26 The addition of MeOH to the water samples was also investigated, with the aim of minimising adsorption of analytes to the vial surface. MeOH was added at five levels from 1% to 30% (v/v). In accordance with the general sample preparation procedure, 1% MeOH was considered as the reference value. As shown in Fig. 2B, the extraction efficiencies of analytes decreased proportionally with increasing addition of the organic solvent, and evidently any extra addition of MeOH was ineffective.
Fig. 2 Influence of addition of salt NaCl (0–10% (w/v)) (A) and organic solvent MeOH (B) on extraction efficiency for selected analytes (analytes are indicated in bold in Table 1). |
As the extraction efficiency was not enhanced by the addition of NaCl or MeOH, all subsequent SBSE experiments were performed only in the presence of 1% MeOH, which was inevitable as the consequence of spiking water blanks with standards dissolved in methanol during the preparation of test samples.
The influences of extraction time and temperature on extraction efficiency were then tested. Taking into account the initial optimised extraction conditions, samples were extracted at four temperature levels between room temperature (25 °C) and 50 °C. Elevated temperatures were achieved using a multi-position digital magnetic hotplate stirrer (Ika Werke, Germany). The relative increase or decrease in extraction efficiency for each analyte was calculated in relation to its extraction efficiency at room temperature (Fig. 3). The results showed that heating the samples improved the extraction efficiency and the highest temperature tested (50 °C) provided the best results for 12 of the tested analytes or at least comparable results to other tested temperatures; the exception was triclosan, where the extraction efficiency was lowest at 50 °C. Overall, extraction efficiencies were up to 6.5-fold higher at the highest temperature tested. In line with evidence on the stability of the analytes at higher temperatures (data not shown), the temperature of 50 °C was selected as the final optimised temperature for SBSE.
Fig. 3 Influence of the sample temperature on extraction efficiency for the selected analytes (compared with the reference temperature of 25 °C). |
In the next step, the influence of the extraction time on extraction efficiency was evaluated by measuring sample response at regular intervals up to 180 min under predetermined conditions (Fig. 4). Since the maximal extraction efficiency for all analytes was reached after 150 min or less, any further prolongation of extraction time was not considered justified and 150 min was adopted as the most efficient compromise between method sensitivity and time consumption.
Fig. 4 Extraction efficiency as a function of extraction time for selected analytes (analytes are indicated in bold in Table 1). |
Lastly, to determine whether increasing the sample volume improved the method sensitivity, the extraction efficiency with sample volumes between 20 mL and 100 mL was measured (Fig. 5). The increase in the response for each analyte was calculated in relation to the response in 20 mL. Responses obtained by increasing the sample volume did not increase proportionally: using 100 mL of sample during the extraction process increased the relative response no more than 2.5-fold instead of an estimated 5-fold. As the extraction time was set at 1 hour, a complete equilibrium may not have been reached in all samples – especially the ones with higher volumes leading to apparent poor correlation between the sample volume and analyte responses. The fact that a higher sample volume would demand inconveniently long extraction times to reach equilibrium23 and the above-mentioned poor correlation between the sample volume and extraction efficiency in initial experiments, a volume of 20 mL was selected as the optimum sample volume for further optimisation experiments.
Fig. 5 Influence of the sample volume on analyte response (compared with responses in the reference volume of 20 mL). |
In order to enhance desorption, a few desorption solvents were examined (Fig. 6). Desorption conditions were set at 15 min and 50 °C. The tested organic solvents all contained acetonitrile, either solely or in combination with MeOH (50:50 v/v), or with the addition of formic acid (0.2% final concentration) or ammonium formate (2.5 mmol final concentration). All the solvents showed good and sufficient extraction efficiency (calculated as relative extraction efficiency) but the most promising (the best for 10 of 17 tested analytes) was a mixture of ACN and MeOH (50:50 v/v). This was selected as the optimal desorption solvent.
Following determination of the most suitable solvent for desorption, the best overall extraction efficiency was examined. Desorption was performed by stirring or sonication or by their combination (Fig. 7); the desorption time was 15 min with stirring at 50 °C and/or 15 min of sonication at room temperature. There was no solvent replacement as the combination of stirring and sonication was used. The resulting extraction efficiencies showed the superiority of stirring over sonication (calculated as the relative extraction efficiency), and their combination improved the extraction efficiency when compared with sonication alone. In some cases the improvement gained by additional sonication was negligible; therefore, stirring only was selected as the standard liquid desorption protocol for further experiments.
Next, extraction efficiency as a function of desorption time (i.e. desorption profiles) was evaluated between 5 and 20 min at 50 °C. The amount of analyte desorbed significantly increased during the first 5 min, followed by a slight increase between 5 and 15 min; further stirring did not result in any extra increase of analyte desorbed (Fig. 8). Even though the difference in the desorption efficiency between 5 and 15 min of stirring appears negligible, at 15 min the repeatability was improved and therefore 15 min of stirring was selected as optimal for further experiments.
Fig. 8 Extraction efficiency as a function of desorption time for selected analytes (analytes are indicated in bold in Table 1). |
Afterwards, stir bars were subjected to four consecutive desorption processes, each time with a fresh desorption solvent. The contribution of consecutive desorptions was calculated as the absolute or a relative enhancement compared with the extraction efficiency during the first desorption process. The individual relative enhancements were mostly below 5%, cumulatively up to 10% (Table 2); however, such desorption processes are time consuming and therefore additional desorption steps were omitted.
Analyte | Extraction efficiency (%) | |||||||
---|---|---|---|---|---|---|---|---|
1st | 2nda | 3rda | 4tha | Total | ||||
a Relative enhancement compared with first desorption efficiency is shown in parentheses. | ||||||||
Amitriptyline | 89.0 | 4.0 | (4.5) | 2.0 | (2.2) | 1.5 | (1.6) | 96.5 |
Clomipramine | 41.2 | 2.4 | (5.8) | 1.2 | (2.9) | 0.9 | (2.2) | 45.7 |
Desipramine | 64.1 | 1.4 | (2.2) | 0.5 | (0.8) | 0.3 | (0.5) | 66.4 |
Donepezil | 34.4 | 1.1 | (3.3) | 0.6 | (1.7) | 0.4 | (1.3) | 36.5 |
Escitalopram | 67.5 | 3.5 | (5.2) | 2.0 | (2.9) | 1.4 | (2.0) | 74.3 |
Fluoxetine | 38.3 | 1.3 | (3.4) | 0.6 | (1.7) | 0.5 | (1.3) | 40.7 |
Haloperidol | 92.6 | 4.3 | (4.6) | 2.1 | (2.3) | 1.6 | (1.7) | 100.6 |
Loperamide | 33.3 | 0.8 | (2.4) | 0.4 | (1.1) | 0.3 | (0.8) | 34.7 |
Loratadine | 79.6 | 1.7 | (2.2) | 1.0 | (1.2) | 0.7 | (0.8) | 82.9 |
Promethazine | 71.8 | 0.3 | (0.5) | 0.2 | (0.3) | 0.3 | (0.4) | 72.6 |
Propranolol | 17.9 | 0.8 | (4.2) | 0.4 | (2.3) | 0.3 | (1.8) | 19.4 |
Selegiline | 69.9 | 3.5 | (5.0) | 1.5 | (2.2) | 0.7 | (1.0) | 75.5 |
Sertraline | 85.2 | 5.4 | (6.3) | 2.6 | (3.0) | 2.0 | (2.3) | 95.1 |
Triclosan | 43.8 | 6.0 | (13.6) | 3.0 | (6.8) | 1.8 | (4.1) | 54.5 |
Venlafaxine | 22.7 | 0.8 | (3.4) | 0.4 | (1.7) | 0.3 | (1.3) | 24.2 |
Verapamil | 62.6 | 1.1 | (1.7) | 0.5 | (0.8) | 0.4 | (0.6) | 64.6 |
Ziprasidone | 11.5 | 0.2 | (1.9) | 0.0 | (0.4) | 0.1 | (0.6) | 11.9 |
Lastly, the temperature of the desorption process was tested in a range from room temperature (25 °C) to 50 °C, maintaining the previously set optimal parameters (Fig. 9). For each analyte, the relative increase in the overall extraction efficiency was compared with the extraction efficiency at room temperature. As expected, similarly to the results of testing the temperature effects during the extraction process, the higher the desorption temperature the better the extraction efficiency. For example, the relative extraction efficiencies were from 1.2-fold to 1.7-fold higher at 50 °C than at room temperature. Accordingly, a temperature of 50 °C was selected for further experiments.
Fig. 9 Influence of the sample heating on extraction efficiency during the desorption process for selected analytes (compared with responses at the reference temperature of 25 °C). |
Analyte | logKO/Wa | pKaa | Linear range (ng L−1) | R 2 | LOQ (ng L−1) | Biasb (%) | RSD (%) n = 6 | EEc (%) | Theoretical EEd (%) |
---|---|---|---|---|---|---|---|---|---|
a Predicted by ALOGPS 2.1 online software. b Data shown at LOQ. c Data shown at 250 ng L−1. d Calculated using eqn (1); logKO/W, partition coefficient (octanol/water); R2, determination coefficient; LOQ, limit of quantification; RSD, relative standard deviation; EE, extraction efficiency. e Analytes in italics were excluded due to their too-low extraction efficiency. | |||||||||
Amitriptyline | 5.10 | 9.76 | 1.25–1250 | 0.9996 | 1.25 | −1.0 | 3.3 | 113 | 100 |
Azithromycin | 3.03 | 9.57 | 1.25–500 | 0.9425 | 1.25 | −18.0 | 20.2 | 24 | 73 |
Desipramine | 4.02 | 10.02 | 1.25–500 | 0.9994 | 1.25 | −3.7 | 1.5 | 101 | 96 |
Diazepam | 2.63 | 2.92 | 1.25–500 | 0.9990 | 1.25 | 0.0 | 4.3 | 53 | 52 |
Donepezil | 4.14 | 8.62 | 1.25–1250 | 0.9939 | 1.25 | −17.8 | 0.9 | 92 | 97 |
Escitalopram | 3.58 | 9.78 | 1.25–1250 | 0.9950 | 1.25 | −6.4 | 1.7 | 81 | 90 |
Fluoxetine | 4.09 | 9.80 | 1.25–1250 | 0.9995 | 1.25 | −7.1 | 1.3 | 122 | 97 |
Haloperidol | 3.70 | 8.05 | 1.25–1250 | 0.9953 | 1.25 | −6.3 | 2.3 | 83 | 93 |
Imatinib | 3.47 | 8.27 | 5.00–1250 | 0.9678 | 5.00 | −7.0 | 13.2 | 7 | 88 |
Clomipramine | 5.04 | 9.20 | 1.25–1250 | 0.9992 | 1.25 | −0.3 | 2.9 | 116 | 100 |
Loperamide | 4.44 | 9.41 | 1.25–500 | 0.9960 | 1.25 | −2.8 | 4.3 | 65 | 99 |
Loratadine | 4.80 | 4.33 | 1.25–1250 | 0.9993 | 1.25 | −6.8 | 3.2 | 100 | 99 |
Metoprolol | 1.80 | 9.67 | 1.25–500 | 0.9914 | 1.25 | −18.7 | 3.3 | 7 | 14 |
Promethazine | 4.52 | 9.05 | 1.25–1250 | 0.9990 | 1.25 | −1.5 | 2.4 | 108 | 99 |
Propranolol | 3.03 | 9.67 | 1.25–500 | 0.9958 | 1.25 | −9.6 | 2.1 | 55 | 73 |
Raloxifene | 5.45 | 7.95 | 5.00–1250 | 0.9959 | 5.00 | −7.0 | 13.3 | 8 | 100 |
Risperidone | 3.27 | 8.76 | 1.25–500 | 0.9981 | 1.25 | −5.0 | 5.8 | 14 | 82 |
Selegiline | 3.08 | 8.67 | 1.25–1250 | 0.9996 | 1.25 | −6.0 | 2.1 | 90 | 75 |
Sertraline | 5.06 | 9.85 | 1.25–1250 | 0.9996 | 1.25 | −7.4 | 2.4 | 130 | 100 |
Tramadol | 2.71 | 9.23 | 1.25–500 | 0.9971 | 1.25 | −8.5 | 4.9 | 49 | 56 |
Triclosan | 5.53 | 7.68 | 1.25–1250 | 0.9988 | 1.25 | −18.7 | 11.6 | 45 | 100 |
Venlafaxine | 2.69 | 8.91 | 1.25–500 | 0.9939 | 1.25 | −0.9 | 3.6 | 60 | 55 |
Verapamil | 5.23 | 9.68 | 1.25–1250 | 0.9996 | 1.25 | −1.0 | 1.8 | 115 | 100 |
Ziprasidone | 4.42 | 7.22 | 1.25–500 | 0.9967 | 1.25 | −7.2 | 5.3 | 18 | 100 |
Bromazepam | 2.09 | 2.68 | 5.00–1250 | 0.6603 | 5.00 | −9.2 | 8.2 | 2 | 24 |
Carbamazepine | 2.10 | −3.80 | 5.00–500 | 0.9976 | 5.00 | −2.3 | 2.7 | 2 | 24 |
Clonazepam | 2.76 | 1.86 | 5.00–500 | 0.9948 | 5.00 | −4.2 | 2.0 | 1 | 59 |
Analyte | Frequency (%) | Concentration (ng L−1) | ||||
---|---|---|---|---|---|---|
A | B | C | D | E | ||
a nd, not detected; LOQ, limit of quantification. | ||||||
Amitriptyline | 100 | 29.0 | 17.5 | 7.8 | 22.0 | 9.8 |
Azithromycin | 100 | 5290.0 | 320.2 | 188.1 | 5271.0 | 508.3 |
Desipramine | 40 | nd | nd | 5.5 | 6.0 | nd |
Diazepam | 60 | nd | 37.8 | 24.0 | 12.2 | nd |
Donepezil | 40 | nd | nd | 20.0 | 17.5 | nd |
Escitalopram | 100 | 54.1 | 186.5 | 89.2 | 92.7 | 72.4 |
Fluoxetine | 40 | nd | 46.7 | 21.8 | nd | nd |
Haloperidol | 100 | 1.3 | 11.1 | 10.1 | 4.2 | 5.0 |
Loperamide | 60 | nd | 4.7 | 4.4 | 3.6 | nd |
Loratadine | 100 | 7.2 | 21.7 | 10.6 | 8.6 | 6.1 |
Metoprolol | 100 | 50.8 | 68.2 | 27.1 | 31.3 | 80.8 |
Promethazine | 20 | nd | nd | 52.2 | nd | nd |
Propranolol | 100 | 55.7 | 62.8 | 18.9 | 13.9 | 32.8 |
Raloxifene | 80 | <LOQ | nd | <LOQ | <LOQ | <LOQ |
Risperidone | 80 | 1.3 | 2.1 | 1.8 | 3.3 | nd |
Selegiline | 40 | nd | 6.2 | 2.4 | nd | nd |
Sertraline | 100 | 6.1 | 2.9 | 1.7 | 174.0 | 2.7 |
Tramadol | 100 | 26106.8 | 25964.0 | 6123.2 | 8605.5 | 10560.2 |
Triclosan | 100 | 53.1 | 126.8 | 71.1 | 280.0 | 39.1 |
Venlafaxine | 100 | 111.2 | 324.5 | 157.8 | 186.5 | 169.7 |
Verapamil | 100 | 15.8 | 29.5 | 13.4 | 23.8 | 12.2 |
Analyte | Concentration (ng L−1) | ||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
WWTP I | WWTP II | WWTP III | WWTP IV | WWTP V | |||||||||||
In | Eff | Removal (%) | In | Eff | Removal (%) | In | Eff | Removal (%) | In | Eff | Removal (%) | In | Eff | Removal (%) | |
a nd, not detected; —, not applicable; In, influent sample; Eff, effluent sample; WWTP, wastewater treatment plant; LOQ, limit of quantification. | |||||||||||||||
Amitriptyline | 17.3 | 15.4 | 10.6 | 6.3 | 5.6 | 11.5 | 17.9 | 8.7 | 51.6 | 18.0 | 5.4 | 70.1 | 6.9 | 7.6 | −10.1 |
Azithromycin | 146.3 | nd | 100.0 | 434.8 | 276.8 | 36.3 | 3029.9 | 103.9 | 96.6 | 156.0 | 137.6 | 11.8 | 171.0 | 153.4 | 10.3 |
Desipramine | <LOQ | nd | 100.0 | 3.2 | <LOQ | — | 1598.1 | 3.8 | 99.8 | 4.9 | 4.3 | 12.4 | 3.0 | 2.3 | 20.6 |
Diazepam | 53.4 | 12.9 | 75.8 | 20.3 | 17.3 | 14.6 | 8.2 | 2.7 | 67.1 | 25.1 | 16.0 | 36.1 | 22.1 | 22.3 | −1.0 |
Donepezil | 10.7 | 8.9 | 17.0 | 18.3 | 11.5 | 37.1 | 10.9 | 7.1 | 34.7 | 16.9 | 14.7 | 13.2 | 11.1 | 10.1 | 8.4 |
Escitalopram | 110.7 | 96.7 | 12.6 | 148.1 | 143.5 | 3.0 | 64.3 | 44.1 | 31.5 | 71.6 | 61.4 | 14.2 | 71.0 | 69.1 | 2.6 |
Fluoxetine | 14.5 | 8.1 | 44.0 | nd | nd | — | 7.6 | nd | — | nd | nd | — | 17.6 | nd | 100.0 |
Haloperidol | 13.5 | 8.1 | 40.3 | 57.5 | 4.6 | 92.0 | 19.4 | nd | — | 5.9 | 4.5 | 23.3 | 11.5 | 6.7 | 41.2 |
Clomipramine | <LOQ | nd | — | nd | nd | — | 0.1 | nd | — | nd | nd | — | <LOQ | 1.7 | — |
Loperamide | 11.0 | <LOQ | — | <LOQ | <LOQ | — | nd | nd | — | 3.6 | <LOQ | — | <LOQ | 2.1 | — |
Loratadine | 2.6 | 3.3 | −23.5 | 9.7 | 7.5 | 22.6 | 3.2 | 2.2 | 30.5 | 10.2 | 8.4 | 17.2 | 6.3 | 6.5 | −2.3 |
Metoprolol | 80.7 | 62.5 | 22.6 | 45.4 | 25.2 | 44.5 | 85.3 | 23.2 | 72.7 | 24.5 | 13.0 | 46.8 | 48.3 | 44.6 | 7.7 |
Propranolol | 64.2 | 47.1 | 26.6 | 40.3 | 39.3 | 2.4 | 42.6 | 11.0 | 74.2 | 45.5 | 19.8 | 56.4 | 45.5 | 42.3 | 6.9 |
Promethazine | nd | nd | — | nd | nd | — | nd | nd | — | nd | nd | — | 12.6 | nd | — |
Raloxifene | <LOQ | <LOQ | — | nd | nd | — | nd | nd | — | <LOQ | <LOQ | — | <LOQ | <LOQ | — |
Risperidone | 2.7 | 2.6 | 4.8 | 5.1 | <LOQ | — | 3.5 | nd | 100.0 | 2.7 | <LOQ | — | 6.7 | nd | — |
Selegiline | 5.6 | 4.7 | 17.5 | nd | nd | — | 3.0 | <LOQ | — | nd | nd | — | nd | nd | — |
Sertraline | 509.0 | 160.0 | 68.6 | 169.7 | 166.4 | 1.9 | 2.5 | 1.7 | 30.9 | 4.6 | 1.9 | 58.1 | 2.9 | 2.5 | 15.4 |
Tramadol | 9488.8 | 13910.4 | −46.6 | 18737.2 | 43860.2 | −134.1 | 6599.6 | 15239.6 | −130.9 | 11392.0 | 6359.9 | 44.2 | 3230.9 | 8618.1 | −166.7 |
Triclosan | 325.3 | 61.9 | 81.0 | 47.1 | 11.7 | 75.2 | 106.8 | 38.8 | 63.7 | 3483.0 | 45.0 | 98.7 | 137.7 | 133.8 | 2.8 |
Venlafaxine | 345.2 | 354.0 | −2.5 | 274.3 | 550.6 | −100.7 | 254.3 | 144.8 | 43.1 | 232.0 | 202.3 | 12.8 | 85.0 | 109.0 | −28.3 |
Verapamil | 29.4 | 26.3 | 10.5 | 64.1 | 49.1 | 23.4 | 13.7 | 1.6 | 88.5 | 12.2 | 12.0 | 1.9 | 21.3 | 20.9 | 1.7 |
The main focus of our paper is not to discuss the removal efficiency results but rather to emphasize the applicability of the method to a variety of purposes. According to the data, tramadol was measured in the highest concentrations and showed extreme resistance to the removal processes. Its effluent concentrations in four out of five WWTP samples actually increased when compared with influent concentrations. Similar results were obtained for venlafaxine. However, it should be noted that analyte concentrations may fluctuate significantly over time and that results may be highly dependent on the sampling technique. Numerous factors can significantly affect pollutant concentrations; for example, precipitations (rain) and hydraulic retention time of wastewater in a WWTP.
Such irregularities are minimised as far as possible in 24 h composite sampling, where concentrations measured reflect the best estimations of actual conditions, when on-line measurements are not available. Overall, removal efficiencies varied from 2% to complete (100%) elimination of the monitored pharmaceuticals present in the influent samples (Table 5).
A brief overlook found that the best removal efficiency was achieved in WWTP III, which is not surprising as it employs a moving-bed biofilm reactor (membrane bioreactor), whereas the other sampled WWTPs used fixed-bed treatments (conventional active sludge technology). For the pharmaceutical pollutants that were monitored, it could be claimed that biological treatment, either conventional or advanced, has a key role in terms of removal efficiency, since there was no significant difference between secondary (biological nutrient removal) and tertiary treatment (disinfection, mainly UV light) (Table 6).
WWTP | Population served (PE) | Flow (×1000 m3 per year) | Technology | Removal efficiencya (%) average; median; range |
---|---|---|---|---|
a Data indicating induction of particular analyte during removal process (negative removal rate) were excluded; PE, population equivalent. | ||||
I | 4000 | 237 | Secondary | 42.1; 26.6; 4.8–100 |
II | 4000 | 585 | Secondary | 30.4; 23.0; 1.9–75 |
III | 55000 | 1790 | Tertiary | 63.2; 65.4; 30.5–100 |
IV | 50000 | 7093 | Tertiary | 34.5; 23.3; 1.9–98.7 |
V | 60000 | 4702 | Tertiary | 19.8; 8.4; 1.7–100 |
This journal is © The Royal Society of Chemistry 2017 |