Issue 8, 2016

Ultra-trace metallic element detection in liquid samples using laser induced breakdown spectroscopy based on matrix conversion and crosslinked PVA polymer membrane

Abstract

A novel method with high sensitivity was developed for determining ultra-trace-amounts of metallic elements in liquids by LIBS analysis, based on crosslinked polyvinyl alcohol (PVA) supporting material. In this newly proposed approach, an aqueous solution sample was intensively mixed with PVA solution leading to phase conversion, which is distinguished from the desiccation process on a solid supporter as reported in the literature of liquid-to-solid conversion for LIBS analysis. The analytes were distributed in a thin PVA film (0.34 mm) as a result of crosslinking the polymer materials, and thereby the plasma signals generated by laser pulse were stronger and a lower amount of solid supporter was needed compared to traditional methods. Due to the low PVA thermal decomposition temperature, laser generated plasma energy was effectively utilized for not only ablating the sample, but rather exciting analytes within the sample. The limits of detection for Ag, Cu, Cr, Pb, Ni, Co and Cd obtained in this study were 1 ng mL−1, 8 ng mL−1, 16 ng mL−1, 1 μg mL−1, 1 μg mL−1, 2 μg mL−1 and 5 μg mL−1, respectively, which showed great improvement compared with those from direct liquid analysis by LIBS. The satisfactory figures of merits of the proposed method, including linearity, reproducibility and accuracy, were also obtained with the optimal experimental parameters. This unique method takes advantage of mixing an aqueous solution sample with supporter material in a liquid phase and it has characteristics of fast, simple and sensitive analysis for liquid sample compared with conventional liquid-to-solid conversion methods for LIBS.

Graphical abstract: Ultra-trace metallic element detection in liquid samples using laser induced breakdown spectroscopy based on matrix conversion and crosslinked PVA polymer membrane

Article information

Article type
Paper
Submitted
09 May 2016
Accepted
25 May 2016
First published
01 Jun 2016

J. Anal. At. Spectrom., 2016,31, 1622-1630

Ultra-trace metallic element detection in liquid samples using laser induced breakdown spectroscopy based on matrix conversion and crosslinked PVA polymer membrane

Q. Lin, X. Han, J. Wang, Z. Wei, K. Liu and Y. Duan, J. Anal. At. Spectrom., 2016, 31, 1622 DOI: 10.1039/C6JA00177G

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements