Racheli
Ron
a,
David
Zbaida
a,
Ilan Z.
Kafka
b,
Rita
Rosentsveig
a,
Ilan
Leibovitch
bc and
Reshef
Tenne
*a
aDepartment of Materials and Interfaces, Weizmann Institute, 234 Herzl street, Rehovot 76100, Israel
bDepartment of Urology, Meir Medical Center, 59 Tchernichovsky street, Kfar Sava, 44410, Israel
cSackler Faculty of Medicine, Tel-Aviv University, Tel-Aviv, Israel. E-mail: reshef.tenne@weizmann.ac.il; Tel: +972-8-9342394 Tel: +972-54-6244242
First published on 20th February 2014
Ureteral stents and urethral catheters are commonly used medical devices for maintaining urinary flow. However, long-term placement (>30 days) of these devices in the urinary tracts is limited by the development of encrustation, a phenomenon that holds a prevalence of 50% within this patient population, resulting in a great deal of morbidity to the patients. Here we report the influence of surface coating of an all-silicone catheter with rhenium-doped fullerene-like molybdenum disulfide (Re:IF-MoS2) nanoparticles on the growth and attachment of in vitro encrustation stones. Scanning electron microscopy (SEM), energy dispersive X-ray spectroscopy (EDS), X-ray photoelectron spectroscopy (XPS) and X-ray powder diffraction (XRD) analyses indicated a remarkable attenuation in encrustation occupation on the Re:IF-MoS2-coated catheter surfaces compared to neat catheters. The doped nanoparticles displayed a unique tendency to self-assemble into mosaic-like arrangements, modifying the surface to be encrustation-repellent. The mechanism of encrustation retardation on the surface coated catheters is discussed in some detail. The ramification of these results for the clogging of other body indwelling devices is briefly discussed.
Encrustation on catheters and stents results from precipitation of ionic salts, which are kept aqueous in the urinal medium under normal conditions.4 Common crystalline deposits on catheters are calcium and magnesium phosphates in the form of hydroxyapatite Ca10(PO4)6(OH)2 and struvite (NH4MgPO4·6H2O).1–3 Other calcium-phosphate phases, like brushite (CaHPO4·2H2O), were also detected on catheter encrustations.6
A great variety of efforts to counteract the problem of encrustation have been reported, using many different strategies and approaches.1 Nonetheless, so far the investigation has failed to eliminate it altogether.
The present work reports the mitigation of encrustation of a commercial all-silicone catheter by a surface coating of inorganic fullerene-like MoS2 (IF-MoS2) nanoparticles.
Nanoparticles of inorganic layered compounds, such as WS2 and MoS2, are known to form a closed-cage fullerene-like structure. These nanoparticles were first reported in 199210,11 and were discussed extensively in several review papers.12,13 An SEM micrograph of a Re:IF-MoS2 powder is shown in Fig. S1A.† The structure and shape of an individual IF-MoS2 nanoparticle is shown by the transmission electron microscopy (TEM) micrograph in Fig. S1B.† Each polyhedral multilayer nanoparticle is made of closed MoS2 layers, and its size is in the range of 50–200 nm. Since an IF-MoS2 nanoparticle is a seamless closed-cage moiety, no structured-edges comprising plenty of dangling bonds are attendant. Hence, each nanoparticle is enveloped by van der Waals surfaces of fully saturated bonded (sulfur-terminated) atoms, possessing a low surface energy of 20 meV Å−2.14 Therefore, the nanoparticles exhibit very low affinity to their environment and they can easily roll or slide.15 Their unique atomically smooth topology (other than a few defects) and the fully saturated-bonds on their surfaces confer these nanoparticles (IF-WS2, IF-MoS2) superior solid-lubrication behavior,16–18 which has been recently exploited commercially.
Doping of the semiconducting IF-MoS2 nanoparticles with foreign atoms of rhenium (<0.015 at%) endows them with n-type conductivity by substitution of molybdenum atoms in the lattice. It is believed that the free electrons produced by the substantial rhenium doping are trapped by defects on the nanoparticle surface; so this extra charge is trapped at the surface or nearby. The Re-doped nanoparticles (Re:IF-MoS2) become negatively charged, thus repelling each other at close proximity. The rhenium doping was shown to have a profound effect on their physio-chemical behavior, rendering them a superior solid-lubricant.19 It was rationalized therefore that films of the electron rich Re:IF-MoS2 nanoparticles together with their superior solid-lubrication behavior16–18 may attenuate encrustation development on catheter surfaces.
In the current study, a commercial all-silicone catheter was coated with a film of Re:IF-MoS2 nanoparticles. A joint incubation of uncoated and Re:IF-MoS2-coated catheter specimens in an in vitro model of a catheterized urinary tract under encrustation conditions was followed. The nature and degree of the encrustation deposits which were developed on the uncoated and Re:IF-MoS2-coated catheter surfaces were comparatively assessed.
Herein, the interactions between encrustation concretions with Re:IF-MoS2 nanoparticles coating are studied in vitro. There exists no single standard for an in vitro model simulating encrustation. Thus, a variety of models for such experiments are reported in the literature.20 The use of artificially made urine provides a very convenient way for comparative studies of catheters coated with different surface functional groups. Moreover, human urine differs between one donor to another, and even between different micturitions of a single donor, thus favouring the use of a standard artificial urine source.
It is shown herein that the presence of a thin film of the Re:IF-MoS2 nanoparticles on the surfaces of the prosthetic device leads to a substantial attenuation in the encrustation on the catheter surface. It is believed that their atomically smooth, passivated-surface and negative surface charge of the nanoparticles delegate the device surface with low drag and adhesive characteristics, minimizing thereby the encrustation phenomena on the catheter surfaces. Hence, the Re:IF-MoS2 film coating would be expected to increase the indwelling durability and decrease associated morbidities to the patient.
Fig. 3 shows SEM micrographs of both uncoated (Fig. 3A and B) and Re:IF-MoS2-coated (Fig. 3C and D) catheter surfaces after an in vitro encrustation process (see also Fig. S2†). The SEM micrographs are colored to facilitate a rapid discrimination between encrustation precipitates and the coating's nanoparticles. The insets of Fig. 3A and C display the original (without coloration) SEM micrographs for comparison. The SEM examination showed that, on both surface types, the solid precipitates were composed of two major morphologies: spherical precipitate with a perforated and poorly crystalline structure and elongated needle-like crystals. In accordance with studies using similar artificial urine,21 the globular precipitates are reminiscent of hydroxyapatite21 and the elongated crystals are assigned as brushite.
Indeed, chemical analysis of the encrusted surfaces by EDS analyses confirmed the growth of calcium- and phosphorus-containing stones. Fig. S4† exhibits the EDS spectra (accompanied by the respective SEM micrograph) which were generated from the two major typical morphologies of the grown deposits. Table 1 quantitatively summarizes the EDS results. A small magnesium quantity was observed in the spectra of the globularly shaped deposits. Nevertheless, struvite (MgNH4PO4·6H2O) crystals were not observed in these series of experiments, most likely due to the low-basic urine pH in the current experimentation, which in most cases did not exceed 7.5.21 The experimentally observed Mg atoms are believed to be incorporated in the poorly crystalline apatite precipitate. The Ca/P ratio in both the elongated crystals and the structureless stones was close to 1. This ratio is typical of brushite whereas it deviates markedly from the composition of (fully crystallized) hydroxyapatite (1.6). As is further detailed below, comparatively XRD and XPS analyses of the surfaces of uncoated and Re:IF-MoS2-coated catheter specimens also confirmed presentation of calcium-phosphate stones.
Structural EDS results [at%] | |||||||
---|---|---|---|---|---|---|---|
Morphology | C (K) | O (K) | Si (K) | P (K) | Ca (K) | Mg (K) | Ca/P |
Brushite | 11.8 | 65.2 | 0.7 | 11.9 | 10.4 | — | 0.9 |
Hydroxyapatite | 32.3 | 46.8 | 5.5 | 7.1 | 7.2 | 1.0 | 1.0 |
Time-dependent experiments showed that the amount of encrustation deposits increased with elongated incubation times for either the uncoated or Re:IF-MoS2-coated catheters. Longer incubation periods led to creation of thicker compact calcium phosphate films on the uncoated catheters. On the other hand, in the case of the Re:IF-MoS2-coated samples, new nuclei were not observed to be formed to a noticeable amount, but an enlargement of the already existing, sporadically distributed, calcium-phosphate precipitates was observed with longer incubation time, mainly in the vertical direction (perpendicular to the substrate).
Using backscattering electron (BSE) imaging mode an enlarged surface area (32000 μm2) over the encrusted Re:IF-MoS2-coated catheter specimen could be carefully analyzed (see Materials and methods). The BSE mode was utilized since the deposited nanoparticles and encrusted stones were barely distinguishable using the SE detector at low magnifications (which could allow the examination of larger surface areas on each sample) (compare Fig. 4A and B). As seen in Fig. 4B, an evenly gray appearance represents the crust in the BSE-generated micrograph, which is much darker than the brighter gray-level of the nanoparticles. Fig. 4C–F exhibit verification of the fidelity of the much faster BSE analyses by a preceding EDS-elemental mapping of the same area. Each (rather long scan-time) EDS-map highlights the surface distribution of a single element of this sample. The congruence between the locations of stones as reflected from the Ca- and P-EDS maps and the BSE signal is clear. Thus, the BSE analysis revealed that the portion of the encrusted area over the enlarged surface (32000 μm2) on the Re:IF-MoS2-coated catheter was 0.44% compared to 92.9% over a similar area of the uncoated catheter.
Table 2 presents an elemental-quantification by EDS of uncoated and Re:IF-MoS2-coated catheter surfaces after an encrustation process. The EDS analysis also found a considerable diminution of encrustation on the encrusted Re:IF-MoS2-coated specimen. Specifically, the Ca and P contents were 10.1 and 8.6 at% (respectively) for the neat specimen against 1.2 and 1.9 at% for the Re:IF-MoS2-coated one. This result is consistent with the SE and BSE analyses.
EDS Encrustation quantification [at%] | ||||||||
---|---|---|---|---|---|---|---|---|
Specimen | C (K) | O (K) | Si (K) | Mo (L) | S (K) | P (K) | Ca (K) | Mg (K) |
Pristine catheter | 35.9 | 31.7 | 32.4 | 0.0 | 0.0 | 0.0 | 0.0 | 0.0 |
Encrusted uncoated catheter | 26.9 | 37.3 | 16.5 | 0.0 | 0.0 | 8.6 | 10.1 | 0.6 |
Encrusted IF-coated catheter | 28.9 | 20.5 | 19.2 | 10.7 | 17.0 | 1.9 | 1.2 | 0.6 |
A couple of catheter specimens were additionally studied by XPS. Fig. 5A presents the broad-scans of encrusted, Re:IF-MoS2-coated and uncoated, catheters, together with the highly resolved binding energy peaks of Ca(2p) and P(2p). Detailed atomic percent composition of the catheter surfaces achieved by the XPS measurements is presented in Table 3. The pronounced difference in the contents of encrustation-related elements is clearly displayed by the XPS results: the atomic concentrations of Ca(2p) and P(2p) for the uncoated catheter were 0.74 and 0.99 at%, respectively. The concentration of these atoms was reduced to 0.21 and 0.15 at% for the Re:IF-MoS2-coated catheter. As expected, the atomic percent ratio of Mo to S on the Re:IF-MoS2-coated specimen was 0.5.
XPS results [at%] | |||||||||
---|---|---|---|---|---|---|---|---|---|
Catheter surface | Ca (2p) | P (2p) | Si (2p) | O (1s) | C (1s) | S (2s) | S (2p) | Mo (3d) | |
Non-encrusted specimens | Uncoated | — | — | 24.46 | 23.69 | 51.48 | — | — | — |
Re:IF-MoS2-coated | — | — | 20.30 | 20.47 | 43.28 | 6.21 | 6.49 | 3.13 | |
Encrusted specimens | Uncoated | 0.74 | 0.99 | 22.75 | 24.26 | 51.25 | — | — | — |
Re:IF-MoS2-coated | 0.21 | 0.15 | 22.17 | 23.07 | 48.83 | 2.13 | 2.31 | 1.15 |
XRD measurements (Fig. 5B) of the encrusted uncoated specimen identified two intense peaks with 2θ values of 11.6 and 23.4 degrees, which can be assigned to the calcium-phosphate phase of brushite (black spectrum in Fig. 5B). In accordance with previous works,21,22 the poorly crystalline hydroxyapatite could not be identified in this pattern. The XRD pattern of the encrusted Re:IF-MoS2-coated catheter exhibited the MoS2 peaks, while the brushite phase had largely vanished, confirming thereby the level of encrustation on this substrate which was under the detectable amount (0.5 wt%).
The experimental results principally demonstrate that the self-assembled Re:IF-MoS2 nanoparticle film has a clear attenuating effect on the encrustation of all-silicon catheters.
Stones of similar types were detected on both the uncoated and Re:IF-MoS2-coated catheter samples. Therefore, the presence of the Re:IF-MoS2 nanoparticles on the coated catheter specimens influenced neither the morphology nor the chemical composition of the in vitro encrustation. However, the difference in the degree of encrustation was consistently detected regardless of the technique used for the analysis.
The exact mechanism of the encrustation suppression on Re:IF-MoS2-coated catheters is not fully comprehensible, yet. Nevertheless, a few key physio-chemical properties of these nanoparticles might provide guidelines for this mechanism. Particularly, their charge, low surface free energy and nano-texture are unique properties which are delegated to the coated catheter surface. Therefore, the presence of the nanoparticle film on the catheter surface alters its nanostructure, as well as its chemistry, influencing thereby the physio-chemical characteristics of the catheter surface.
Two different encrustation mechanisms can be considered; one of them involves a direct nucleation of the hydroxyapatite and brushite at stable surface-sites enabling their further growth on the available area. Simultaneously, stones nucleate and grow in the solution and subsequently these colloidal nanoparticles precipitate/adhere on the catheter surface.
The observed massive supersaturation in the urine during encrustation experiments indicates the enormous amount of colloidal stones surrounding each incubated catheter specimen. This situation introduces an abundant possibility of stone precipitation and adherence. However, as investigation of the surface-structure of the Re:IF-MoS2-coated catheters showed, the colloidal stone particles approaching a catheter specimen from the bulk solution encounter an architecture totally different from the smooth substrate of a neat catheter specimen. Generation of the special surface-nanostructure by self-assembly of the negatively charged Re:IF-MoS2 nanoparticles into two-dimensional close-packed arrays (Fig. 1C, 2 and S3A†) produced a compact dense array of nano-distant bumps. This nanoparticle arrangement might reduce the availability of contact-points for anchoring stones onto the Re:IF-MoS2-coated catheter surface, which is an established mechanism for self-cleaning surfaces.23–25
Additionally, due to its atomically smooth surface and low surface energy,14 the Re:IF-MoS2 material is known to be chemically very inert26,27 and induce very low friction.16–18 Therefore, the anchoring potential of the hydroxyapatite colloidal nanoparticles to the underlying substrate is expected to be very low. Consequently, the stones are believed to “slip” on the Re:IF-MoS2-coated catheters once approaching the surface (Fig. 6).
The visual absence of encrustation on bald areas on the Re:IF-MoS2-coated catheter surfaces suggests that, even if some encrustation has occurred on these areas, these patches could be easily uprooted by the dynamic flow of the urine.
The specific structure and chemistry of each Re:IF-MoS2 nanoparticle provides further support to the above model; the low surface energy (20 meV Å−2)14 of the basal (0001) 2H-MoS2 surface implies that the terminal (sulphur) atoms are very inert with respect to a specific chemical reaction under the present conditions.26,27 However, the curved (0001) surfaces of the IF-MoS2 nanoparticles contain a small amount (<5%) of structural defects, which are chemically reactive, but can be passivated via adsorption of specific moieties.28 Such defects can be the source of the rather rare and random growth of stones on the surface coated catheters.
Future experiments are aimed at investigating the influence of such Re:IF-MoS2 nanoparticles coatings on indwelling medical devices of different sorts, like blood stents, or metallic implants. The biocompatibilities of the IF-MoS2 nanoparticles are currently being investigated. Early indications suggest that these nanoparticles are benign and do not pose any toxic risk.29
A suspension of 0.05 wt% Re:IF-MoS2 nanoparticles in ultrapure H2O (Milli-Q RG, Millipore) was sonicated, using an ultra-sonic probe mixer (Vibra Cell VCX400, 400 W, Sonics & Materials) for 30 min. The ultra-sonication was alternately applied (6 s activation and 4 s deactivation) on the Re:IF-MoS2-suspension, during which a constant magnetic stirring was implemented.
Catheter specimens were individually suspended in vials containing 10 ml of the Re:IF-MoS2 suspension. The vials were left 24 hours for mixing using a rotation machine. Prior to the analyses, the catheter specimens were removed from the Re:IF-MoS2-suspension and were rinsed with ultrapure H2O. Moreover, for the sake of comparison, uncoated bare catheter specimens were put inside similar vials which contained 10 ml H2O, and were treated through the same procedure. Additional series of samples were prepared from pristine silicone catheters for reference purposes. These samples were analyzed without any prior treatment.
In the present work, the encrustation processes were performed in a glass reaction vessel equipped with a fitting lid within which 12 marked stainless steel rods were equally positioned. At the end of each rod stood a hook on which a single vertical specimen was positioned. The vessel was placed in an incubator to maintain the physiological temperature (37 °C). In order to systematically reproduce the conditions of the encrustation process along the experiments, an artificial urine solution was used. This solution has a well-defined composition, compared to the human urine which has a non-uniform composition among different native donors and within different micturitions of an individual donor. The artificial urine solution used in the present work was similar to that used by Griffith et al.30 The solution consisted of 10 solutes (see Table S1,† initial pH = 6.4–6.5), whose concentrations were equivalent to the average concentration found over a 24 hour period in the urine of normal human.30 Alkalinization of the urinal medium was triggered here by a direct addition of a Jackbean-derived urease (type III, Sigma-Aldrich). The urease was supplied as a lyophilized powder, which was dissolved in a pre-prepared filtered (0.45 μm) sodium phosphate buffer (2 M, pH = 7.0). In most of the reported experiments, the urease concentrations and incubation times were approx. 0.05 mg per 100 ml (5 ppm) urine solution and approximately 8–12 hours, respectively. The enzyme powder was found to be very hygroscopic, affecting the actual enzyme solution concentrations. Therefore, it is important to emphasize that the incubation period, i.e. the time lapse for turbidity, varied somewhat from one series of measurement to the other, depending on the freshness of the enzyme powder which is very hygroscopic. Thus, care was taken to maintain the enzyme under strictly dry conditions. Furthermore, during all sets of experiments, the two kinds of specimens (Re:IF-MoS2-coated and uncoated) were simultaneously incubated in the encrustation reactor. The samples were taken out for analysis as soon as the solution lost its full transparency and became massively turbid. Upon removal of the encrusted specimens out of the in vitro model, they were gently rinsed with ultrapure H2O to remove loosely attached debris and were stored inside an evacuated desiccator till further analyses. Furthermore, during each experiment pH measurements were carried out by a pH-meter (pH510, Eutech Instruments).
EDS (EDAX instrument, Phoenix, attached to the E-SEM) was used for the chemical analysis of the specimens. Here, two modes of work were implemented for the chemical analysis. In the first one, the beam was focused on a single stone (high magnification). The analysis was repeated three times for stones of the same morphology and the result is reported as an average of the three measurements. In addition, a global (low magnification) EDS analysis of the encrusted surface (3000 μm2) was carried out. The EDS analyses of bare, encrusted, uncoated and Re:IF-MoS2-coated catheter specimens were compared. All the low magnification EDS analyses were performed by sampling three distinct surface locations. The results are reported as the average of the three EDS measurements. The accelerating voltage of the beam for the EDS analysis was limited to 15 keV.
SEM imaging and image analysis were used to perform quantitative analysis of the encrustation developed on the catheter surface. Imaging the catheter surface with the SE detector proved to be rather problematic for this purpose. Discrimination between the precipitated stones and the Re:IF-MoS2 nanoparticles was effective at high magnification (×20000), only. However, the heterogeneity of the surface did not permit acquiring sufficient data for a fully quantitative analysis under high magnification. Conversely, at low magnification the discrimination between the stones and the Re:IF-MoS2 nanoparticles was not adequate in the SE mode. Therefore, mapping the encrusted surface with a BSE detector, which is sensitive to the atomic number (Z), combined with image analysis was preferred for the quantitative analysis. The contrast difference in the BSE mode allowed clear discrimination among the encrusted stones, the Re:IF-MoS2 nanoparticles and the substrate at lower magnification (×5000) and over large surface areas, thus enabling quantitative analysis of the different substrates.
Confirmation of the BSE mapping with EDS analysis, which is slow and rather tedious, was done with full agreement between the two analyses. The surface area of a Re:IF-MoS2 nanoparticle-coated specimen after encrustation was analyzed by dividing the surface into a raster (mesh). Each raster unit was scanned by the BSE detector at a magnification of ×5000 (50 micrographs, 640 μm2 each, 32000 μm2 total area). Image analysis of the BSE data was done using the ImageJ (National Institutes of Health, USA) software.
Samples for the SEM analyses were prepared in the following manner: ∼5 × 5 mm2 samples were cut from the middle of each parent sample. A thin layer of gold–palladium was evaporated on each specimen using a high vacuum evaporation set-up (S150 sputter coater, Edwards). For the EDS analysis, carbon evaporation was applied instead, using a high vacuum evaporation set-up (BOC FL400, Edwards).
To minimize the beam damage effects, the analysis time was set as 30 min. Curve fitting analysis was based on linear background subtraction and application of Gaussian–Lorenzian line shapes. Quantification was carried-out using the peak area, and corrected with Scofield sensitivity factors. Signals were collected from an area size of 900 × 400 μm2 for each sample.
In addition to the encrusted specimens, a few non-encrusted samples (prior to urine exposure) for control were also analyzed including: Re:IF-MoS2-coated and also bare untreated catheter specimens.
Footnote |
† Electronic supplementary information (ESI) available. See DOI: 10.1039/c3nr06231g |
This journal is © The Royal Society of Chemistry 2014 |