Issue 5, 2017

Microfluidic platform for rapid measurement of transepithelial water transport

Abstract

Water transport across epithelial monolayers is of central importance in mammalian fluid homeostasis, and epithelial aquaporin (AQP) water channels are potential drug targets. Current methods to measure transepithelial water permeability based on indicator dilution have limited accuracy and can require hours for a single measurement. We report here a microfluidics platform for rapid and accurate measurement of water transport across a conventionally cultured epithelial monolayer on a porous filter requiring only a single image obtained using a standard laboratory fluorescence microscope. The undersurface of a porous polyester filter containing cultured epithelial cells on top is contacted with a perfused microfluidic channel of 100 μm width, 20 μm height and 10 cm length with folded geometry, with in-plane size of 3.2 × 3.2 mm2 for visualization with a 2× objective lens. Osmotic water permeability is measured from the steady-state concentration profile along the length of the channel of a membrane-impermeant fluorescent dye in the perfusate, in which an osmotic gradient is imposed by an anisosmolar solution overlying the epithelial monolayer; diffusional water permeability is measured using a D2O/H2O-sensing fluorescent dye in the perfusate with a D2O-containing isosmolar solution overlying the cell layer. Permeability values are deduced from single fluorescence images. The method, named fluid transport on a chip (FT-on-Chip), was applied to measure transepithelial osmotic and diffusional water permeability in control and AQP4-expressing epithelial cell monolayers. FT-on-Chip allows for rapid, accurate and repeated measurements of transepithelial water permeability, and is generalizable to transport measurements of ions and solutes using suitable indicator dyes.

Graphical abstract: Microfluidic platform for rapid measurement of transepithelial water transport

Supplementary files

Article information

Article type
Paper
Submitted
28 Nov 2016
Accepted
06 Feb 2017
First published
06 Feb 2017

Lab Chip, 2017,17, 887-895

Microfluidic platform for rapid measurement of transepithelial water transport

B. Jin and A. S. Verkman, Lab Chip, 2017, 17, 887 DOI: 10.1039/C6LC01456A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.