Integration of a wet analysis system on a glass chip: determination of Co(II) as 2-nitroso-1-naphthol chelates by solvent extraction and thermal lens microscopy

Author affiliations

Abstract

The integration of a wet analysis system on a glass chip was demonstrated and determination of Co(II) was performed using this system. The Co(II) was extracted into m-xylene from aqueous solution as 2-nitroso-1-naphthol chelates, and colorimetric determination of the m-xylene phase was applied by a thermal lens microscope. The integration of the chemical operation procedures shown here leads to a considerable reduction in analyzing time. The time for extraction in the integrated system, 10 min, was about tenfold shorter than a conventional system using a separatory funnel and mechanical shaker. Moreover, troublesome operations such as phase separation necessary for the conventional system could be omitted. The determination of Co(II) in the range 2 × 10−7–1 × 10−8 M, which was estimated to be 0.072–1.44 zmol, was achieved.

Back to tab navigation
Please wait while Download options loads

Publication details

The article was received on 27 Mar 2001, accepted on 30 May 2001 and first published on 09 Aug 2001


Article type: Paper
DOI: 10.1039/B102790P
Citation: Lab Chip, 2001,1, 72-75