Communication. Assessment of direct solid sample analysis by graphite pellet electrothermal vaporization inductively coupled plasma mass spectrometry

Abstract

Graphite pellet vaporization was evaluated as a method for solid sample introduction for ICP-MS. A detection limit of 0.1 ppb of Cd was obtained indicating that the technique might be very useful for screening. The accuracy was poor using the technique of external standards and four reference materials, suggesting that the technique of standard additions might be required if the technique was to be used for anything other than semi-quantitative measurements.

Article information

Article type
Paper

J. Anal. At. Spectrom., 1995,10, 1027-1029

Communication. Assessment of direct solid sample analysis by graphite pellet electrothermal vaporization inductively coupled plasma mass spectrometry

J. M. Ren, R. Rattray, E. D. Salin and D. C. Grégoire, J. Anal. At. Spectrom., 1995, 10, 1027 DOI: 10.1039/JA9951001027

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.