Issue 3, 2012

Evaluation of the diffusion size classifier (meDiSC) for the real-time measurement of particle size and number concentration of nanoaerosols in the range 20–700 nm

Abstract

In the frame of assessing exposure to nanostructured particles, the aim of this work is to study the performance of a new device devoted to the real-time measurement of nanostructured aerosol: the meDiSC (Diffusion Size Classifier, Matter Engineering, Switzerland). This instrument is based on unipolar diffusion charging of particles which are then collected successively in diffusion and filtration stages. From currents measured in these stages, the instrument is capable of determining aerosol mean size and number concentration. These data were compared to reference measurements regarding monodisperse aerosols in a range from 20 to 700 nm; the relative biases were found unsatisfying. This led us to investigate the principle of the instrument. Consequently, the charging law of the diffusion charger was experimentally established, as well as the calibration curve allowing the determination of the mean size of the particles. The latter analysis was completed by a model based on diffusion theory. Our results indicate the possibility to improve the range of size measurement up to 350 nm. Measured particle size and number concentration were also used to calculate geometric surface-area concentration; the experimental data were compared to a reference calculated surface-area concentration. The results demonstrate the possibility to evaluate this parameter within acceptable uncertainty. In a second step, the meDiSC was challenged with polydisperse aerosols. It was observed that meDiSC overestimates particle size by a factor 1.7, while particle number concentrations are found within ±40% of the reference. The model applied to polydisperse aerosols indicates that polydispersity little influences particle size up to 300 nm while geometric standard deviation remains below 1.7.

Graphical abstract: Evaluation of the diffusion size classifier (meDiSC) for the real-time measurement of particle size and number concentration of nanoaerosols in the range 20–700 nm

Article information

Article type
Paper
Submitted
29 Jul 2011
Accepted
10 Jan 2012
First published
14 Feb 2012

J. Environ. Monit., 2012,14, 1014-1023

Evaluation of the diffusion size classifier (meDiSC) for the real-time measurement of particle size and number concentration of nanoaerosols in the range 20–700 nm

S. Bau, J. Jacoby and O. Witschger, J. Environ. Monit., 2012, 14, 1014 DOI: 10.1039/C2EM10619A

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.