Issue 9, 2003

Ion/molecule reactions performed in a miniature cylindrical ion trap mass spectrometer

Abstract

A recently constructed miniature mass spectrometer, based on a cylindrical ion trap (CIT) mass analyzer, is used to perform ion/molecule reactions in order to improve selectivity for in situ analysis of explosives and chemical warfare agent simulants. Six different reactions are explored, including several of the Eberlin reaction type (M. N. Eberlin and R. G. Cooks, Org. Mass Spectrom., 1993, 28, 679–687) as well as novel gas-phase Meerwein reactions. The reactions include (1) Eberlin transacetalization of the benzoyl, 2,2-dimethyloximinium, and 2,2-dimethylthiooximinium cations with 2,2-dimethyl-1,3-dioxolane to form 2-phenyl-1,3-dioxolanylium cations, 2,2-dimethylamine-1,3-dioxolanylium cations and the 2,2-dimethylamin-1,3-oxathiolanylium cations, respectively; (2) Eberlin reaction of the phosphonium ion CH3P(O)OCH3+, formed from the chemical warfare agent simulant dimethyl methylphosphonate (DMMP), with 1,4-dioxane to yield the 1,3,2-dioxaphospholanium ion, a new characteristic reaction for phosphate ester detection; (3) the novel Meerwein reaction of the ion CH3P(O)OCH3+ with propylene sulfide forming 1,3,2-oxathionylphospholanium ion; (4) the Meerwein reaction of the benzoyl cation with propylene oxide and propylene sulfide to form 4-methyl-2-phenyl-1,3-dioxolane and its thio analog, respectively; (5) ketalization of the benzoyl cation with ethylene glycol to form the 2-phenyl-1,3-dioxolanylium cation; (6) addition/NO2 elimination involving benzonitrile radical cation in reaction with nitrobenzene to form an arylated nitrile, a diagnostic reaction for explosives detection and (7) simple methanol addition to the C7H7+ ion, formed by NO2 loss from the molecular ion of p-nitrotoluene to form an intact adduct. Evidence is provided that these reactions occur to give the products described and their potential analytical utility is discussed.

Article information

Article type
Paper
Submitted
31 Mar 2003
Accepted
27 Jun 2003
First published
29 Jul 2003

Analyst, 2003,128, 1112-1118

Ion/molecule reactions performed in a miniature cylindrical ion trap mass spectrometer

L. S. Riter, E. C. Meurer, E. S. Handberg, B. C. Laughlin, H. Chen, G. E. Patterson, M. N. Eberlin and R. G. Cooks, Analyst, 2003, 128, 1112 DOI: 10.1039/B303609J

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.

Read more about how to correctly acknowledge RSC content.

Social activity

Spotlight

Advertisements