Optimisation in inductively coupled plasma mass spectrometry

Abstract

The performance of a commercial inductively coupled plasma mass spectrometer has been investigated as a function of a number of parameters associated with sample introduction and plasma operation. The forward r.f. power to the plasma and the nebuliser argon pressure are the two that have the greatest effect. For a variety of elements, similar optimum settings for these two variables are found. Optimising the signal due to M+ generally minimises the signal due to M2+ or MO+. Substantial differences between the optima for these variables in ICP-MS compared with ICP-ES are found.

Article information

Article type
Paper

Analyst, 1986,111, 901-906

Optimisation in inductively coupled plasma mass spectrometry

S. E. Long and R. M. Brown, Analyst, 1986, 111, 901 DOI: 10.1039/AN9861100901

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.