Uncertainty in CCD detectors with and without cooling devices when used for molecular fluorescence measurements

Author affiliations

Abstract

The performance of CCD detectors for fluorescence measurements is evaluated through uncertainty studies, mainly as a function of both the signal intensity and the detector temperature. Two CCD detectors, one furnished with a cooling device and the other without a cooling device, were used, and the results were compared. The dependence of uncertainty on the instrumental signal was evaluated simultaneously for both detectors at temperatures ranging between −23 and 23 °C. The tested detectors required temperature increments between 30 and 50 °C to double dark noise (the random part of dark current). The detector temperature affects uncertainty, but this is basically related to dark noise in fluorescence measurements. The consequence of this is that the temperature does not significantly affect the signal-to-noise ratio (S/N); thus, the presence of a cooling device does not significantly improvement the performance, and the limit of detection (LOD) does not depend strongly on the detector temperature. The quality of the two-dimensional array does affect uncertainty, the value of S/N, and the LOD. A good CCD detector can perform at a level similar to a spectrofluorometer furnished with a photomultiplier tube. Laboratory data are given to show how the three components of uncertainty (dark noise, shot noise and flicker noise) behave at different signal intensities and temperatures. Dark noise is the most important factor, shot noise has relevance only at high signal values, and flicker noise is practically irrelevant. The classical model of the dependence of S/N on the fluorescence signal is applied, and the uncertainty constants that govern the performance of the apparatus used are given.

Graphical abstract: Uncertainty in CCD detectors with and without cooling devices when used for molecular fluorescence measurements

Back to tab navigation

Publication details

The article was received on 10 Dec 2014, accepted on 29 Jan 2015 and first published on 30 Jan 2015


Article type: Paper
DOI: 10.1039/C4AY02934H
Author version
available:
Download author version (PDF)
Citation: Anal. Methods, 2015,7, 2379-2385
  •   Request permissions

    Uncertainty in CCD detectors with and without cooling devices when used for molecular fluorescence measurements

    I. Aramburu, J. Galbán, M. Ostra, C. Ubide, M. Vidal and J. Zuriarrain, Anal. Methods, 2015, 7, 2379
    DOI: 10.1039/C4AY02934H

Search articles by author

Spotlight

Advertisements