Jump to main content
Jump to site search

Issue 15, 2014
Previous Article Next Article

A 1024-sample serum analyzer chip for cancer diagnostics

Author affiliations

Abstract

We present a platform that combines microarrays and microfluidic techniques to measure four protein biomarkers in 1024 serum samples for a total of 4096 assays per device. Detection is based on a surface fluorescence sandwich immunoassay with a limit of detection of ~1 pM for most of the proteins measured: PSA, TNF-α, IL-1β, and IL-6. To validate the utility of our platform, we measured these four biomarkers in 20 clinical human serum samples, 10 from prostate cancer patients and 10 female and male controls. We compared the results of our platform to a conventional ELISA and found a good correlation between them. However, compared to a classical ELISA, our device reduces the total cost of reagents by 4 orders of magnitude while increasing throughput by 2 orders of magnitude. Overall, we demonstrate an integrated approach to perform low-cost and rapid quantification of protein biomarkers from over one thousand serum samples. This new high-throughput technology will have a significant impact on disease diagnosis and management.

Graphical abstract: A 1024-sample serum analyzer chip for cancer diagnostics

Back to tab navigation

Supplementary files

Publication details

The article was received on 09 Oct 2013, accepted on 25 Nov 2013 and first published on 26 Nov 2013


Article type: Paper
DOI: 10.1039/C3LC51153G
Author version available: Download Author version (PDF)
Citation: Lab Chip, 2014,14, 2642-2650
  •   Request permissions

    A 1024-sample serum analyzer chip for cancer diagnostics

    J. L. Garcia-Cordero and S. J. Maerkl, Lab Chip, 2014, 14, 2642
    DOI: 10.1039/C3LC51153G