Jump to main content
Jump to site search

Issue 7, 2011
Previous Article Next Article

Polyatomic interferences in plutonium determination in the femtogram range by double-focusing sector-field ICP-MS

Author affiliations

Abstract

The determination of very low concentrations of plutonium, in the femtogram per millilitre range, in environmental samples using ICP-MS can be interfered by polyatomic species containing heavy elements, from hafnium to bismuth. Mass-to-charge ratios of these polyatomic species are very close to the ones of the plutonium isotopes and cannot be separated from them in the low resolution mode currently used for trace analysis. In this paper, we present an evaluation of the extent and impact of these interferences on the background at masses of plutonium isotopes 239Pu and 240Pu, based on measurement of environmental samples for three years using a double-focusing sector-field ICP-MS. We demonstrate that these molecular interferences, especially the ones involving lead, mercury, and iridium through species PbO2+, ArHg+, IrO3+, must be considered as the additional background induced by these interferences (on average for all interferences respectively ∼1.2 and ∼1.9 counts s−1 at 239 and 240 atomic mass units) is often higher than the instrumental background measured with deionised water acidified at 2% by ultra-pure grade HNO3 (∼0.4 counts s−1), and are generally of the same order as other usual sources of background for low-level plutonium analysis by ICP-MS, i.e.242Pu tracer isotopic impurities, 238U hydrides and peak tail. Thus, if not corrected, these polyatomic interferences may lead to false detection of femtogram amounts of plutonium or overestimation of results. The extents of formation of the main polyatomic species are highly variable from one analysis to the other, over one or two orders of magnitude, depending on instrumental settings. Chemical purification procedures are very efficient in eliminating the greater part of heavy elements present in the initial samples. However, these procedures also bring heavy elements into the final sample solutions through contamination by the atmosphere, glassware and reagents.

Graphical abstract: Polyatomic interferences in plutonium determination in the femtogram range by double-focusing sector-field ICP-MS

Back to tab navigation

Publication details

The article was received on 30 Jul 2010, accepted on 07 Mar 2011 and first published on 08 Apr 2011


Article type: Paper
DOI: 10.1039/C0JA00097C
Citation: J. Anal. At. Spectrom., 2011,26, 1474-1480
  •   Request permissions

    Polyatomic interferences in plutonium determination in the femtogram range by double-focusing sector-field ICP-MS

    F. Pointurier, A. Hubert, A. Fauré, P. Hémet and A. Pottin, J. Anal. At. Spectrom., 2011, 26, 1474
    DOI: 10.1039/C0JA00097C

Search articles by author

Spotlight

Advertisements