On-chip three dimensional microcoils for MRI at the microscale

Author affiliations

Abstract

We present for the first time a fully MEMS-integrated technology to manufacture 3D geometrically perfect solenoidal microcoils for microscale MRI applications. We report 25 µm isotropic resolution MR images of a copper sulfate aqueous phantom. These images are acquired using microcoils with 5 windings of insulated 25 µm diameter Au wire and with quality factors as high as 46 at the operating frequency (400 MHz).

Graphical abstract: On-chip three dimensional microcoils for MRI at the microscale

Back to tab navigation
Please wait while Download options loads

Publication details

The article was received on 13 Jan 2010, accepted on 26 Mar 2010 and first published on 21 Apr 2010


Article type: Communication
DOI: 10.1039/C000840K
Citation: Lab Chip, 2010,10, 1387-1390
  •   Request permissions

    On-chip three dimensional microcoils for MRI at the microscale

    V. Badilita, K. Kratt, N. Baxan, M. Mohmmadzadeh, T. Burger, H. Weber, D. v. Elverfeldt, J. Hennig, J. G. Korvink and U. Wallrabe, Lab Chip, 2010, 10, 1387