Tunable external cavity quantum cascade laser for the simultaneous determination of glucose and lactate in aqueous phase

Author affiliations

Abstract

A room temperature operated pulsed external-cavity (EC) quantum cascade laser (QCL) was used for mid-infrared (mid-IR) transmission measurements of glucose and lactate in aqueous solution. The high spectral power density of the EC-QCL (ranging from 1–350 mW) over a wide tuning range (1030–1230 cm−1) allowed transmission measurements through optical paths of 130 μm and more. This is a significant improvement in terms of robustness of the measurement setup, especially when samples containing cells or other particles, as is the case for biofluids, are to be analyzed. The broad tuning range furthermore permitted multi-analyte detection based on multivariate calibrations. Promising results on the simultaneous determination of glucose (c = 0–800 mg dL−1) and sodium-lactate (c = 0–224 mg dL−1) in aqueous solutions in the presence of the interferents maltose and xylose are reported. A partial least squares (PLS) calibration model was calculated which was able to predict the glucose concentration with a root mean square error of prediction (RMSEP) of 9.4 mg dL−1, as proved by external validation. Due to their small size and room temperature operation, EC-QCLs offer an attractive alternative regarding the way mid-IR measurements are carried out. This may be of special importance for new reagent-free bedside monitoring systems.

Graphical abstract: Tunable external cavity quantum cascade laser for the simultaneous determination of glucose and lactate in aqueous phase

Back to tab navigation
Please wait while Download options loads

Publication details

The article was received on 16 Jul 2010, accepted on 11 Oct 2010 and first published on 02 Nov 2010


Article type: Paper
DOI: 10.1039/C0AN00532K
Citation: Analyst, 2010,135, 3260-3265
  •   Request permissions

    Tunable external cavity quantum cascade laser for the simultaneous determination of glucose and lactate in aqueous phase

    M. Brandstetter, A. Genner, K. Anic and B. Lendl, Analyst, 2010, 135, 3260
    DOI: 10.1039/C0AN00532K

Search articles by author