High-sensitivity miniaturized immunoassays for tumor necrosis factor α using microfluidic systems

Author affiliations

Abstract

We use microfluidic chips to detect the biologically important cytokine tumor necrosis factor α (TNF-α) with picomolar sensitivity using sub-microliter volumes of samples and reagents. The chips comprise a number of independent capillary systems (CSs), each of which is composed of a filling port, an appended microchannel, and a capillary pump. Each CS fills spontaneously by capillary forces and includes a self-regulating mechanism that prevents adventitious drainage of the microchannels. Thus, interactive control of the flow in each CS is easily achieved via collective control of the evaporation in all CSs by means of two Peltier elements that can independently heat and cool. Long incubation times are crucial for high sensitivity assays and can be conveniently obtained by adjusting the evaporation rate to have low flow rates of ∼30 nL min−1. The assay is a sandwich fluorescence immunoassay and takes place on the surface of a poly(dimethylsiloxane) (PDMS) slab placed across the microchannels. We precoat PDMS with capture antibodies (Abs), localize the capture of analyte molecules using a chip, then bind the captured analyte molecules with fluorescently-tagged detection Abs using a second chip. The assay results in a mosaic of fluorescence signals on the PDMS surface which are measured using a fluorescence scanner. We show that PDMS is a compatible material for high sensitivity fluorescence assays, provided that detection antibodies with long excitation wavelength fluorophores (≥580 nm) are employed. The chip design, long incubation times, proper choice of fluorophores, and optimization of the detection Ab concentration all combine to achieve high-sensitivity assays. This is exemplified by an experiment with 170 assay sites, occupying an area of ∼0.6 mm2 on PDMS to detect TNF-α in 600 nL of a dendritic cell (DC) culture medium with a sensitivity of ∼20 pg mL−1 (1.14 pM).

Back to tab navigation
Please wait while Download options loads

Publication details

The article was received on 15 Jun 2004, accepted on 05 Oct 2004 and first published on 10 Nov 2004


Article type: Paper
DOI: 10.1039/B408964B
Citation: Lab Chip, 2004,4, 563-569
  •   Request permissions

    High-sensitivity miniaturized immunoassays for tumor necrosis factor α using microfluidic systems

    S. Cesaro-Tadic, G. Dernick, D. Juncker, G. Buurman, H. Kropshofer, B. Michel, C. Fattinger and E. Delamarche, Lab Chip, 2004, 4, 563
    DOI: 10.1039/B408964B

Search articles by author