Abstract

Simple criteria and procedures to assess which part of a potentiometric titration curve can be used to obtain an appreciably linear Gran plot and an accurate extrapolation to the equivalence volume are discussed for one-step precipitation, complexometric, redox and acid-base titrations. Equations were derived for calculating the limiting values of the experimental parameters that prevent titration errors higher than an assigned value.

Article information

Article type
Paper

Analyst, 1989,114, 463-470

Linearity range of Gran plots for the end-point in potentiometric titrations

C. Maccà and G. G. Bombi, Analyst, 1989, 114, 463 DOI: 10.1039/AN9891400463

To request permission to reproduce material from this article, please go to the Copyright Clearance Center request page.

If you are an author contributing to an RSC publication, you do not need to request permission provided correct acknowledgement is given.

If you are the author of this article, you do not need to request permission to reproduce figures and diagrams provided correct acknowledgement is given. If you want to reproduce the whole article in a third-party publication (excluding your thesis/dissertation for which permission is not required) please go to the Copyright Clearance Center request page.