Continuous colorimetric monitoring of vapour-phase urea and cyanates

Abstract

A method is described for continuously monitoring the concentrations of urea and cyanates in gases, based on a modification of a colorimetric method used for urea in human serum. The method was calibrated with standard solutions of urea and potassium cyanate. The reproducibility was ±2% throughout. Using this method, the vapour pressure of urea was measured from 58 to 113 °C. The vapour pressure results have an estimated accuracy of 10% in the range 3 × 10–4 to 5 × 10–2 mmHg.

Back to tab navigation

Article type: Paper
DOI: 10.1039/AN9780300759
Citation: Analyst, 1978,103, 759-765
  •   Request permissions

    Continuous colorimetric monitoring of vapour-phase urea and cyanates

    L. E. Trimble and R. J. H. Voorhoeve, Analyst, 1978, 103, 759
    DOI: 10.1039/AN9780300759